Differences between the wear couples metal-on-polyethylene and ceramic-on-ceramic in the stability against dislocation of total hip replacement

被引:34
作者
Bader, R [1 ]
Steinhauser, E
Zimmermann, S
Mittelmeier, W
Scholz, R
Busch, R
机构
[1] Tech Univ Munich, Klin Orthopadie & Sportorthopadie, D-8000 Munich, Germany
[2] Univ Leipzig, Klin & Poliklin Orthopadie, Leipzig, Germany
[3] Tech Univ Munich, Inst Med Stat & Epidemiol, D-8000 Munich, Germany
关键词
D O I
10.1023/B:JMSM.0000030214.79180.13
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
After total hip replacement an insufficient range of motion (ROM) can lead to contact between femoral neck and rim of the cup (= impingement) causing dislocation and consecutive material failure. The purpose of this study was to analyse the influence of different wear couples on the ROM and stability against dislocation. By means of a special testing device the ROM until impingement, the ROM until dislocation as well as the resisting moment against levering the head out of the cup were experimentally determined. Various total hip systems with cup inserts made of ceramic and polyethylene were comparatively examined in different implant positions. Maximum resisting moment as well as the ROM until impingement and dislocation were clearly influenced by the implant position. Furthermore, the stability against dislocation was affected by design parameters, whereas in the case of appropriate implant position differing wear couples (metal-on-polyethylene vs. ceramic-on-ceramic) had a minor impact. However, as shown by tests under lubricant conditions, ceramic-on-ceramic couples provided less dislocation stability in unfavourable implant position in comparison to metal-on-polyethylene. Therefore, ceramic-on-ceramic couples should only be applied in the case of optimised implant orientation preventing impingement and dislocation with subsequent material failure like chipping off or breakage. (C) 2004 Kluwer Academic Publishers.
引用
收藏
页码:711 / 718
页数:8
相关论文
共 25 条
[1]
[Anonymous], 2001, HIP INT
[2]
BADER R, 2003, ORTHOPADISCHE PRAXIS, V39, P430
[3]
Impingement and rim wear associated with early osteolysis after a total hip replacement - A case report [J].
Barrack, RL ;
Schmalzried, TP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (07) :1218-1220
[4]
BEAVER RJ, 2002, BIOCERAMICS JOINT AR, P65
[5]
BERGMANN G, 1994, J BIOMECH, V6, P969
[6]
Ceramic-on-ceramic bearings in total hip arthroplasty [J].
Bierbaum, BE ;
Nairus, J ;
Kuesis, D ;
Morrison, JC ;
Ward, D .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (405) :158-163
[7]
New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty [J].
D'Antonio, J ;
Capello, W ;
Manley, M ;
Bierbaum, B .
JOURNAL OF ARTHROPLASTY, 2002, 17 (04) :390-397
[8]
GARINO JP, 2003, BIOCERAMICS JOINT AR, P215
[9]
Long-term registration has improved the quality of hip replacement - A review of the Swedish THR Register comparing 160,000 cases [J].
Herberts, P ;
Malchau, H .
ACTA ORTHOPAEDICA SCANDINAVICA, 2000, 71 (02) :111-121
[10]
Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration [J].
Kennedy, JG ;
Rogers, WB ;
Soffe, KE ;
Sullivan, RJ ;
Griffen, DG ;
Sheehan, LJ .
JOURNAL OF ARTHROPLASTY, 1998, 13 (05) :530-534