Quantitation of platelet adhesion to Ti and DLC-coated Ti in vitro using 125I-labeled platelets

被引:36
作者
Krishnan, LK
Varghese, N
Muraleedharan, CV
Bhuvaneshwar, GS
Derangère, F
Sampeur, Y
Suryanarayanan, R
机构
[1] Univ Paris 11, Lab Physicochim Etat Solide, UMR8648, F-91405 Orsay, France
[2] ICC, F-77350 Le Mee Sur Seine, France
[3] St Chitra Tirunal Inst Med Sci, Thiruvanandhapuram 690512, India
来源
BIOMOLECULAR ENGINEERING | 2002年 / 19卷 / 2-6期
关键词
diamond like carbon coating; Ti; blood compatibility; radio active scintigraphy; shear stress;
D O I
10.1016/S1389-0344(02)00029-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Measurement of platelet adhesion in vitro is a good indicator of its reactivity to implant devices in vivo. Platelets were labeled with I-125 without affecting its normal morphology and function and the labeled platelets were suspended in platelet poor plasma and exposed to Ti and diamond like carbon-coated (DLC) Ti discs, under dynamic conditions, using a parallel plate flow chamber. The test materials were washed, dried, exposed to a phosphor screen and scanned to get the images. The number of platelets that adhered to Ti was found to be higher than those that adhered to DLC coated Ti sample, irrespective of the shear stress which was varied between 2 and 16 dynes/cm(2). (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:251 / 253
页数:3
相关论文
共 9 条
[1]  
[Anonymous], BIOL EVALUATION MED
[2]  
Dion I., 1993, Bio-Medical Materials and Engineering, V3, P51
[3]   VASCULAR GRAFT THROMBUS FORMATION [J].
HANSON, SR ;
HARKER, LA .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1987, 516 :653-661
[4]   IN-111 (IN-111) AND CHROMIUM-51 (CR-51) LABELING OF PLATELETS - ARE THEY COMPARABLE [J].
HEATON, WAL .
TRANSFUSION, 1986, 26 (01) :16-19
[5]  
Jones MI, 2000, J BIOMED MATER RES, V52, P413, DOI 10.1002/1097-4636(200011)52:2<413::AID-JBM23>3.0.CO
[6]  
2-U
[7]   Diamond-like carbon - present status [J].
Lifshitz, Y .
DIAMOND AND RELATED MATERIALS, 1999, 8 (8-9) :1659-1676
[8]   CORROSION-RESISTANCE OF AMORPHOUS HYDROGENATED SIC AND DIAMOND-LIKE COATINGS DEPOSITED BY RF-PLASMA-ENHANCED CHEMICAL-VAPOR-DEPOSITION [J].
SELLA, C ;
LECOEUR, J ;
SAMPEUR, Y ;
CATANIA, P .
SURFACE & COATINGS TECHNOLOGY, 1993, 60 (1-3) :577-583
[9]  
YANQI Y, 1996, CELLS MAT, V6, P339