Rapid measurement of fish freshness indices by an amperometric flow-injection system with a 16-way switching valve and immobilized enzyme reactors

被引:21
作者
Nanjyo, Y [1 ]
Yao, T [1 ]
机构
[1] Univ Osaka Prefecture, Grad Sch Engn, Dept Appl Chem, Osaka 5998531, Japan
关键词
fish freshness; freshness index; flow-injection analysis; enzyme reactors; amperometric detection; 16-way switching valve;
D O I
10.1016/S0003-2670(02)00717-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel flow-injection system is proposed for the rapid measurement of the fish freshness indices K-1 and K-2: K-1 = [([HxR + [Hx]) x 100/([IMP] + [HxR] + [Hxl)] and K-2 = [[Hx] x 100/([HxR] + [Hx])], where [IMP], [HxR] and [Hx] are inosine-5'-monophosphate, inosine and hypoxanthine concentrations, respectively.: For the estimation of index K1, 5'-nucleotidase immobilized reactor and nucleoside phosphorylase (NP)/xanthine oxidase (XO) coimmobilized reactor were incorporated in series in the flow-injection line made up by 4 16-way switching valve with two sample loops. For the estimation of index K-2, NP and XO immobilized reactors were also incorporated in the similar flow-line. Two sample portions passed through the flow-line with different residence times so that two peaks were obtained. The first and second peaks obtained in the K-1-determining system corresponded to the total of HxR and Hx and the total of Hx, HxR and IMP, respectively. Similarly, the first and second peaks obtained in the K-2-determining system corresponded to Hx and the total of Hx and HxR, respectively. Therefore, the indices K-1 and K-2 can be estimated by K-1 and K-2 (%) = (i(1)/i(2)) x f x 100 where i(1) and i(2) present the peak current of the first and second peaks, respectively, and f the ratio of the peak currents of the first and second peaks for a Hx standard solution. A sea bream was selected as a model fish and it was stored at 4degreesC after death. A good correlation was found between the index K-1 and the storage time over a period of 400 It, with a correlation coefficient of 0.992, but no correlation between the index K-2 and the storage time. The measurements could be performed at a rate of 22 samples per hour with satisfactory precision (0.6-1.2% R.S.D.), without calibration for each species, accurate weighing of fish meat and any interferences in fish meat. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:175 / 183
页数:9
相关论文
共 19 条
[1]   Amperometric detection of uric acid and hypoxanthine with Xanthine oxidase immobilized and carbon based screen-printed electrode. Application for fish freshness determination [J].
Carsol, MA ;
Volpe, G ;
Mascini, M .
TALANTA, 1997, 44 (11) :2151-2159
[2]   Development of a system with enzyme reactors for the determination of fish freshness [J].
Carsol, MA ;
Mascini, M .
TALANTA, 1998, 47 (02) :335-342
[3]  
EHIRA S, 1970, NIPPON SUISAN GAKK, V36, P491
[4]   SIMULTANEOUS HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ASSAY OF THE ACTIVITIES OF ERYTHROCYTIC HYPOXANTHINE GUANINE PHOSPHORIBOSYL TRANSFERASE AND PURINE NUCLEOSIDE PHOSPHORYLASE [J].
HALFPENNY, AP ;
BROWN, PR .
JOURNAL OF CHROMATOGRAPHY, 1985, 349 (02) :275-282
[5]  
Hayashi K, 1996, ANAL LETT, V29, P2499
[6]   Amperometric sensor for fish freshness based on immobilized multi-enzyme modified electrode [J].
Hu, SS ;
Liu, CC .
ELECTROANALYSIS, 1997, 9 (16) :1229-1233
[7]  
KITED Y, 1983, J FOOD HYG SOC JPN, V24, P225
[8]  
LEE EH, 1982, B JPN SOC SCI FISH, V48, P225
[9]  
LOUNG JHT, 1992, ENZYME MICROB TECHNO, V14, P125
[10]   DEVELOPMENT OF A SYSTEM WITH DOUBLE ENZYME REACTORS FOR THE DETERMINATION OF FISH FRESHNESS [J].
OKUMA, H ;
TAKAHASHI, H ;
YAZAWA, S ;
SEKIMUKAI, S ;
WATANABE, E .
ANALYTICA CHIMICA ACTA, 1992, 260 (01) :93-98