Optimization and assessment of quantitative 124I imaging on a Philips Gemini dual GS PET/CT system

被引:27
作者
Gregory, Rebecca A. [1 ,2 ]
Hooker, Claire A. [1 ,2 ]
Partridge, Mike [1 ,2 ]
Flux, Glenn D. [1 ,2 ]
机构
[1] Inst Canc Res, Joint Dept Phys, Sutton SM2 5PT, Surrey, England
[2] Royal Marsden NHS Fdn Trust, Sutton SM2 5PT, Surrey, England
关键词
Iodine-124; Positron-emission tomography; Gadolinium oxyorthosilicate; POSITRON-EMISSION-TOMOGRAPHY; DIFFERENTIATED THYROID-CANCER; RADIOIODINE THERAPY; MONOCLONAL-ANTIBODY; DOSIMETRY; TUMOR; RADIOIMMUNOTHERAPY; I-131; IMPACT;
D O I
10.1007/s00259-009-1099-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Quantitative I-124 PET imaging is challenging as I-124 has a complex decay scheme. In this study the performance of a Philips Gemini dual GS PET/CT system was optimized and assessed for I-124. The energy window giving the maximum noise equivalent count rate (NECR) and NEMA 2001-NU2 image quality were measured. The activity concentration (AC) accuracy of images calibrated using factors from F-18 and I-124 decaying source measurements were investigated. The energy window 455-588 keV gave the maximum NECR of 9.67 kcps for 233 MBq. I-124 and F-18 image quality was comparable, although I-124 background variability was increased. The average underestimation in AC in I-124 images was 17.9 +/- 2.9% for nonuniform background and 14.7 +/- 2.9% for single scatter simulation (SSS) subtraction scatter correction. At 224 MBq the underestimation was 10.8 +/- 11.3%, which is comparable to 7.7 +/- 5.3% for F-18, but increased with decreasing activity. The best I-124 PET quantitative accuracy was achieved for the optimized energy window, using SSS scatter correction and calibration factors from decaying I-124 source measurements. The quantitative accuracy for I-124 was comparable to that for F-18 at high activities of 224 MBq but diminishing with decreasing activity. Specific corrections for prompt gamma-photons may further improve the quantitative accuracy.
引用
收藏
页码:1037 / 1048
页数:12
相关论文
共 31 条
[1]   Optimization of a fully 3D single scatter simulation algorithm for 3D PET [J].
Accorsi, R ;
Adam, LE ;
Werner, ME ;
Karp, JS .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12) :2577-2598
[2]  
[Anonymous], 2001, SCI REPORT, P1
[3]  
Blasberg RG, 2000, CANCER RES, V60, P624
[4]   124I in PET imaging:: impact on quantification, radiopharmaceutical development and distribution [J].
Bockisch, Andreas ;
Freudenberg, Lutz ;
Rosenbaum, Sandra ;
Jentzen, Walter .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 (11) :1247-1248
[5]  
Collingridge DR, 2002, CANCER RES, V62, P5912
[6]  
DAGHIGHIAN F, 1993, EUR J NUCL MED, V20, P402
[7]   MBP-annexin V radiolabeled directly with iodine-124 can be used to image apoptosis in vivo using PET [J].
Dekker, B ;
Keen, H ;
Lyons, S ;
Disley, L ;
Hastings, D ;
Reader, A ;
Ottewell, P ;
Watson, A ;
Zweit, J .
NUCLEAR MEDICINE AND BIOLOGY, 2005, 32 (03) :241-252
[8]  
Erdi Yusuf E., 1999, Clin Positron Imaging, V2, P41, DOI 10.1016/S1095-0397(99)00004-7
[9]   Evaluation of dosimetry of radioiodine therapy in benign and malignant thyroid disorders by means of iodine-124 and PET [J].
Eschmann, SM ;
Reischl, G ;
Bilger, K ;
Kupferschläger, J ;
Thelen, MH ;
Dohmen, BM ;
Besenfelder, H ;
Bares, R .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (06) :760-767
[10]   RADIATION-DOSE ASSESSMENT IN RADIOIODINE THERAPY .2. PRACTICAL IMPLEMENTATION USING QUANTITATIVE SCANNING AND PET, WITH INITIAL RESULTS ON THYROID-CARCINOMA [J].
FLOWER, MA ;
SCHLESINGER, T ;
HINTON, PJ ;
ADAM, I ;
MASOOMI, AM ;
ELBELLI, MA ;
OTT, RJ ;
MCCREADY, VR ;
HARMER, CL .
RADIOTHERAPY AND ONCOLOGY, 1989, 15 (04) :345-357