The identification of external factors which influence the calibration of Coriolis massflow meters

被引:34
作者
Cheesewright, R [1 ]
Clark, C [1 ]
Bisset, D [1 ]
机构
[1] Brunel Univ, Dept Syst Engn, Uxbridge UB8 3PH, Middx, England
关键词
Coriolis meter; calibration error; condition monitoring;
D O I
10.1016/S0955-5986(99)00023-0
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
It has been suggested that, by monitoring the sensor signals and the meter drive signal, it may be possible to create a Coriolis meter which can identify the presence of a variety of external disturbances and can give a warning and/or make a correction to the meter output in order to maintain accuracy. This paper presents the findings from an investigation of the response of commercially available Coriolis meters to such disturbances. The results of this program of tests are reported and the implications of these results on the practicality of the creation of an 'intelligent' meter are discussed. (C) 2000 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1 / 10
页数:10
相关论文
共 13 条
[11]  
HURREN PJ, 1988, P 2 INT C FLOW MEAS
[12]  
NICHOLSON S, 1994, P 7 INT C FLOW MEAS
[13]   EFFECT OF PULSATING FLOW ON CORIOLIS MASS FLOWMETERS [J].
VETTER, G ;
NOTZON, S .
FLOW MEASUREMENT AND INSTRUMENTATION, 1994, 5 (04) :263-273