A Low-Noise CMOS Receiver Frontend for MRI

被引:11
作者
Anders, Jens [1 ]
Boero, Giovanni [1 ]
机构
[1] Ecole Polytech Fed Lausanne, Lausanne, Switzerland
来源
2008 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE - INTELLIGENT BIOMEDICAL SYSTEMS (BIOCAS) | 2008年
关键词
D O I
10.1109/BIOCAS.2008.4696900
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
In this paper a novel architecture for an integrated receiver front-end for micro magnetic resonance imaging (micro-MRI) applications is described. While the chip consumes only 9mA supply current (4mA in the LNA and 5mA in the output buffer) from a 33V power supply, it has a measured input referred noise density of only 0.6 nV/root Hz. The receiver consists of a reception coil, an on-chip tuning capacitor, a low-noise amplifier, and a 50 Omega output buffer. The system is designed for operation in a B(0)-field of 7 Gamma corresponding to a frequency of 300MHz. It is implemented in a 0.35 mu m CMOS high-voltage process and occupies a chip area of 850 mu m x 500 mu m.
引用
收藏
页码:165 / 168
页数:4
相关论文
共 13 条
[1]  
[Anonymous], 2006, ANALOG DESIGN ESSENT
[2]   Fully integrated probe for proton nuclear magnetic resonance magnetometry [J].
Boero, G ;
Frounchi, J ;
Furrer, B ;
Besse, PA ;
Popovic, RS .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (06) :2764-2768
[3]  
Carrington A., 1979, INTRO MAGNETIC RESON
[4]  
CHERIFI T, 2005, IEEE T CIRCUITS SY 1, V52
[5]  
CHO ZH, 1982, P IEEE, V70
[6]  
ENZ C, 1978, EPFL
[7]  
GATTA F, 2001, IEEE J SOLID STATE C, V36
[8]  
Hoult DI, 2000, CONCEPT MAGNETIC RES, V12, P173, DOI 10.1002/1099-0534(2000)12:4<173::AID-CMR1>3.0.CO
[9]  
2-Q
[10]  
KARANICOLAS AN, 1996, IEEE J SOLID STATE C, V31