Optimization of selectivity and resolution in micellar electrokinetic capillary chromatography with a mixed micellar system of sodium dodecyl sulfate and sodium cholate

被引:24
作者
Wiedmer, SK
Jumppanen, JH
Haario, H
Riekkola, ML
机构
[1] UNIV HELSINKI,DEPT CHEM,ANALYT CHEM LAB,FIN-00014 HELSINKI,FINLAND
[2] FINNSUGAR LTD,FINNSUGAR DEV,KANTVIK,FINLAND
[3] UNIV HELSINKI,DEPT MATH,FIN-00014 HELSINKI,FINLAND
关键词
micellar electrokinetic capillary chromatography; mixed micellar system; corticosteroids; optimization; 1,3-diaminopropane;
D O I
10.1002/elps.1150171221
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Selectivity and resolution were studied for the separation of seven corticosteroids by micellar electrokinetic capillary chromatography (MEKC) using a mixed micellar solution of sodium dodecyl sulfate (SDS) and sodium cholate (SC), buffered with 3-(N-morpholino)propanesulfonic acid (MOPS) or 3-[(1,1-dimethyl-2-hydroxyethyl)amino]-2-hydroxypropane sulfonic acid (AMPSO). The changes in selectivity were compared for the AMPSO-SDS-SC system by varying the pH and the concentrations of AMPSO, SDS and SC. The experimental design started with the central composite design and continued in a sequential manner. The optimum selectivity for the separation of the corticosteroids was calculated from the analyte migration times and the analyte velocities, by using empirical quadratic regression models. Satisfactory regression fits and coefficients of determination for prediction were obtained with cross-validated models. To optimize the resolution, the physical parameters of capillary length and analysis time were varied under the conditions optimal for the selectivity. In both the selectivity and the resolution, optimization the overall optimum was determined by using the desirability function technique. Analysis times were controlled by using 1,3-diaminopropane to influence the electroosmotic flow velocity (v(eo)). The voltage was kept constant, which resulted in higher electric field strength in shorter capillaries. No changes in the selectivity were observed when 1,3-diaminopropane was used to control the electroosmotic flow velocity. Such an optimization technique, where the chemical and physical factors affecting the separation are treated independently, seemed to be effective for finding the best possible resolution for the corticosteroids.
引用
收藏
页码:1931 / 1937
页数:7
相关论文
共 29 条
[1]   FREE SOLUTION CAPILLARY ELECTROPHORESIS OF BASIC-PROTEINS IN UNCOATED FUSED-SILICA CAPILLARY TUBING [J].
BULLOCK, JA ;
YUAN, LC .
JOURNAL OF MICROCOLUMN SEPARATIONS, 1991, 3 (03) :241-248
[2]   MIXED MICELLAR ELECTROKINETIC CHROMATOGRAPHY OF CORTICOSTEROIDS [J].
BUMGARNER, JG ;
KHALEDI, MG .
ELECTROPHORESIS, 1994, 15 (10) :1260-1266
[3]  
COLE RO, 1990, HRC-J HIGH RES CHROM, V13, P579
[4]   NUCLEAR MAGNETIC-RESONANCE AND X-RAY STUDIES ON MICELLAR AGGREGATES OF SODIUM DEOXYCHOLATE [J].
CONTE, G ;
DIBLASI, R ;
GIGLIO, E ;
PARRETTA, A ;
PAVEL, NV .
JOURNAL OF PHYSICAL CHEMISTRY, 1984, 88 (23) :5720-5724
[5]   MULTIPLE-CRITERIA OPTIMIZATION [J].
DEMING, SN .
JOURNAL OF CHROMATOGRAPHY, 1991, 550 (1-2) :15-25
[6]   SIZE AND SHAPE OF SODIUM DEOXYCHOLATE MICELLAR AGGREGATES [J].
ESPOSITO, G ;
GIGLIO, E ;
PAVEL, NV ;
ZANOBI, A .
JOURNAL OF PHYSICAL CHEMISTRY, 1987, 91 (02) :356-362
[7]   HYDROGEN-ION BUFFERS FOR BIOLOGICAL-RESEARCH [J].
FERGUSON, WJ ;
BRAUNSCHWEIGER, KI ;
BRAUNSCHWEIGER, WR ;
SMITH, JR ;
MCCORMICK, JJ ;
WASMANN, CC ;
JARVIS, NP ;
BELL, DH ;
GOOD, NE .
ANALYTICAL BIOCHEMISTRY, 1980, 104 (02) :300-310
[8]  
Good N E, 1972, Methods Enzymol, V24, P53
[9]   HYDROGEN ION BUFFERS FOR BIOLOGICAL RESEARCH [J].
GOOD, NE ;
WINGET, GD ;
WINTER, W ;
CONNOLLY, TN ;
IZAWA, S ;
SINGH, RMM .
BIOCHEMISTRY, 1966, 5 (02) :467-&
[10]   OPTIMIZATION OF PRECISION AND THROUGHPUT IN MICELLAR ELECTROKINETIC CHROMATOGRAPHY [J].
HAYASHI, Y ;
MATSUDA, R ;
TERABE, S .
CHROMATOGRAPHIA, 1993, 37 (3-4) :149-155