Pulmonary artery thermodilution cardiac output vs. transpulmonary thermodilution cardiac output in two patients with intrathoracic pathology

被引:12
作者
Breukers, RBGE [1 ]
Jansen, JRC [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Intens Care, NL-2300 RC Leiden, Netherlands
关键词
cardiac output; intrathoracic pathology; transpulmonay cardiac output;
D O I
10.1111/j.1399-6576.2004.00375.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
In two adult patients, one with a severe hemorrhage and one with a partial anomalous pulmonary vein, cardiac output (CO) measurements were performed simultaneously by means of the bolus transpulmonary thermodilution technique (COao) and continuous pulmonary artery thermodilution method (CCOpa). In both cases, the methods revealed clinically significant different cardiac output values based upon the site of measurement and the underlying pathology The assessment of cardiac output (CO) is considered an important part of cardiovascular monitoring of the critically ill patient. Cardiac output is most commonly determined intermittently by the bolus thermodilution technique with a pulmonary artery catheter (COpa). As continuous monitoring of CO is preferable to this intermittent technique, two major techniques have been proposed. Firstly, a nearly continuous thermodilution method (CCOpa) using a heating filament mounted on a pulmonary artery catheter (Baxter Edwards Laboratories, Irvine, CA), with a clinically acceptable accuracy compared with the intermittent bolus technique (1-4). Based on these results we assumed CCOpa equivalent to real CO during hemodynamically stable conditions, and secondly, a continuous cardiac output system based on pulse contour analysis (PCCO), such as the PiCCO system (Pulsion Medical System, Munchen, Germany). To calibrate this device, which uses a derivation of the algorithm of Wesseling and colleagues (5), an independently obtained value of CO by the transpulmonary thermodilution method (COao) is used. Clinical validation studies in patients without underlying intrathoracic pathology, comparing transpulmonary COao with the pulmonary technique (COpa), mostly yielded good agreement (6-10).
引用
收藏
页码:658 / 661
页数:4
相关论文
共 20 条
[1]  
BOECK JC, 1989, J CRIT CARE, V4, P106
[2]  
BOLDT J, 1994, CRIT CARE MED, V22, P1913
[3]  
Bottiger BW, 1996, INTENS CARE MED, V22, P312
[4]   CONTINUOUS VERSUS INTERMITTENT CARDIAC-OUTPUT MEASUREMENT IN CARDIAC SURGICAL PATIENTS UNDERGOING HYPOTHERMIC CARDIOPULMONARY BYPASS [J].
BOTTIGER, BW ;
RAUCH, H ;
BOHRER, H ;
MOTSCH, J ;
SODER, M ;
FLEISCHER, F ;
MARTIN, E .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 1995, 9 (04) :405-411
[5]  
de Figueiredo LFP, 1999, J TRAUMA, V47, P288
[6]   Continuous and intermittent cardiac output measurement:: pulmonary artery catheter versus aortic transpulmonary technique [J].
Della Rocca, G ;
Costa, MG ;
Pompei, L ;
Coccia, C ;
Pietropaoli, P .
BRITISH JOURNAL OF ANAESTHESIA, 2002, 88 (03) :350-356
[7]   Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilution [J].
Goedje, O ;
Hoeke, K ;
Lichtwarck-Aschoff, M ;
Faltchauser, A ;
Lamm, P ;
Reichart, B .
CRITICAL CARE MEDICINE, 1999, 27 (11) :2407-2412
[8]   Continuous cardiac output monitoring during adult liver transplantation: Thermal filament technique versus bolus thermodilution [J].
Greim, CA ;
Roewer, N ;
Thiel, H ;
Laux, G ;
Esch, JSA .
ANESTHESIA AND ANALGESIA, 1997, 85 (03) :483-488
[9]   EVALUATION OF A NEW CONTINUOUS THERMODILUTION CARDIAC-OUTPUT MONITOR IN CRITICALLY ILL PATIENTS - A PROSPECTIVE CRITERION STANDARD STUDY [J].
HALLER, M ;
ZOLLNER, C ;
BRIEGEL, J ;
FORST, H .
CRITICAL CARE MEDICINE, 1995, 23 (05) :860-866
[10]   CONTINUOUS CARDIAC-OUTPUT MEASUREMENTS IN THE PERIOPERATIVE PERIOD [J].
JAKOBSEN, CJ ;
MELSEN, NC ;
ANDRESEN, EB .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1995, 39 (04) :485-488