Scintillation light emission studies of LSO scintillators

被引:33
作者
Saoudi, A [1 ]
Pepin, C
Pépin, C
Houde, D
Lecomte, R
机构
[1] Univ Sherbrooke, Clin Res Ctr, Metab & Funct Imaging Ctr, Sherbrooke, PQ J1K 2R1, Canada
[2] Univ Sherbrooke, Dept Phys, Sherbrooke, PQ J1K 2R1, Canada
[3] Univ Sherbrooke, Dept Nucl Med & Radiobiol, Sherbrooke, PQ J1K 2R1, Canada
关键词
D O I
10.1109/23.819255
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
UV and gamma-ray excited luminescence and nuclear spectroscopy were used to study the relationship between the scintillation mechanisms of LSO and the spectroscopic characteristics obtained with PMT and APD readouts at room temperature. No correlation was found between scintillation decay time and light output. Like other investigators, we observed the existence of two distinct luminescence centers, Ce1 and Ce2, that mainly give rise to short (420 nm) and long (440 nm) emission wavelengths. Our measurements showed that different LSO crystals excited by gamma-rays have emission spectra with largely different shapes and maxima depending on the relative population and luminescence efficiency of these centers. It was also found that the poor energy resolution of LSO and YSO. scintillators is well correlated with the coexistence of the two competing luminescence mechanisms. The prevalence of either Ce1 or Ce2 luminescence tends to reduce the variance of light emission and, thus, to improve energy resolution. Inversely, the coexistence of the two centers increases variance and degrades energy resolution.
引用
收藏
页码:1925 / 1928
页数:4
相关论文
共 21 条
[1]   Investigation of LSO crystals for high spatial resolution positron emission tomography [J].
Casey, ME ;
Eriksson, L ;
Schmand, M ;
Andreaco, MS ;
Paulus, M ;
Dahlbom, M ;
Nutt, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1109-1113
[2]   Performance of a YSO/LSO phoswich detector for use in a PET/SPECT system [J].
Dahlbom, M ;
MacDonald, LR ;
Eriksson, L ;
Paulus, M ;
Andreaco, M ;
Casey, ME ;
Moyers, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1114-1119
[3]   Non-proportionality in the scintillation response and the energy resolution obtainable with scintillation crystals [J].
Dorenbos, P ;
deHaas, JTM ;
vanEijk, CWE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (06) :2190-2202
[4]   NONLINEAR RESPONSE IN THE SCINTILLATION YIELD OF LU2SIO5 CE(3+) [J].
DORENBOS, P ;
DEHAAS, JTM ;
VANEIJK, CWE ;
MELCHER, CL ;
SCHWEITZER, JS .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :735-737
[5]   Investigation of GSO, LSO and YSO scintillators using reverse avalanche photodiodes [J].
Lecomte, R ;
Pepin, C ;
Rouleau, D ;
Saoudi, A ;
Andreaco, MS ;
Casey, M ;
Nutt, P ;
Dautet, H ;
Webb, PP .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :478-482
[6]  
LECOMTE R, 1998, IN PRESS NUCL INST A
[7]   On the scintillation response of NaI(Tl)-crystals [J].
Leutz, H ;
DAmbrosio, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (02) :190-193
[8]   Investigation of a new readout scheme for high resolution scintillation crystal arrays using photodiodes [J].
Levin, CS ;
Hoffman, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1208-1213
[9]   A short-wavelength selective reach-through avalanche photodiode [J].
McIntyre, RJ ;
Webb, PP ;
Dautet, H .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1341-1346
[10]   ROOM-TEMPERATURE LSO PIN PHOTODIODE PET DETECTOR MODULE THAT MEASURES DEPTH OF INTERACTION [J].
MOSES, WW ;
DERENZO, SE ;
MELCHER, CL ;
MANENTE, RA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1085-1089