Investigation of a new readout scheme for high resolution scintillation crystal arrays using photodiodes

被引:11
作者
Levin, CS
Hoffman, EJ
机构
[1] Division of Nuclear Medicine and Biophysics, UCLA School of Medicine, Los Angeles
基金
美国国家卫生研究院;
关键词
D O I
10.1109/23.596989
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We are exploring the possibility of using PIN photodiodes to readout the scintillation crystals used in Positron Emission Tomography (PET) detector designs. Semiconductor photodetectors typically have a lower signal to noise ratio than photomultiplier tubes (PMTs). However, they have the advantage of compactness, and, thus, scintillation crystal readout schemes not available to PMTs because of their size and geometry limitations, are readily available to photodiodes. With current PET detector designs, only a small fraction of the available scintillation light, created from 511 keV gamma ray interactions within the crystal, is collected. Scintillation light collection studies were performed for several crystal geometries and surface treatments using both simulations and measurements. In this report, we present a feasible photodiode readout scheme that allows greater than 90% of the available scintillation light created in either BGO or LSO scintillation crystals to be collected by the photodetector. This improvement in light collection with the photodiode readout somewhat compensates for its lower inherent signal to noise ratios and makes it feasible for use in PET detectors. A coincident timing spectrum resolution of 9.4 ns FWHM was measured for 511 keV interactions with one LSO crystal coupled to a photodiode, the other to a PMT.
引用
收藏
页码:1208 / 1213
页数:6
相关论文
共 19 条
[1]   A HIGH-RESOLUTION DETECTION SYSTEM FOR POSITRON TOMOGRAPHY [J].
BARTON, JB ;
HOFFMAN, EJ ;
IWANCZYK, JS ;
DABROWSKI, AJ ;
KUSMISS, JH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1983, 30 (01) :671-675
[2]   A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY [J].
CASEY, ME ;
NUTT, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :460-463
[3]   A COMPARISON OF PET DETECTOR MODULES EMPLOYING RECTANGULAR AND ROUND PHOTOMULTIPLIER TUBES [J].
CHERRY, SR ;
TORNAI, MP ;
LEVIN, CS ;
SIEGEL, S ;
HOFFMAN, EJ ;
ANDREACO, MS ;
WILLIAMS, CW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1064-1068
[4]   Optical fiber readout of scintillator arrays using a multi-channel PMT: A high resolution PET detector for animal imaging [J].
Cherry, SR ;
Shao, YP ;
Siegel, S ;
Silverman, RW ;
Mumcuoglu, E ;
Meadors, K ;
Phelps, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1932-1937
[5]  
CLINTHORNE NH, 1993, IEEE NUCL SCI S MED, V3, P1692
[6]   AN EVALUATION OF A TWO-DIMENSIONAL ARRAY DETECTOR FOR HIGH-RESOLUTION PET [J].
DAHLBOM, M ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (04) :264-272
[7]   HYBRID MERCURIC IODIDE (HGI2) - GADOLINIUM ORTHOSILICATE (GSO) DETECTOR FOR PET [J].
DAHLBOM, M ;
MANDELKERN, MA ;
HOFFMAN, EJ ;
RICCI, AR ;
BARTON, JB ;
IWANCZYK, JS ;
DABROWSKI, AJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) :533-537
[8]   HIGH-RESOLUTION POSITRON EMISSION TOMOGRAPHY USING SMALL BISMUTH-GERMANATE CRYSTALS AND INDIVIDUAL PHOTOSENSORS [J].
DERENZO, SE ;
BUDINGER, TF ;
VULETICH, T .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1983, 30 (01) :665-670
[9]   GAMMA-RAY SPECTROSCOPY USING SMALL, COOLED BISMUTH-GERMANATE SCINTILLATORS AND SILICON PHOTODIODES [J].
DERENZO, SE .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 219 (01) :117-122
[10]   LIGHT COLLECTION IN SCINTILLATION DETECTOR COMPOSITES FOR NEUTRON DETECTION [J].
KNOLL, GF ;
KNOLL, TF ;
HENDERSON, TM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :872-875