Detection of subtle lung nodules: Relative influence of quantum and anatomic noise on chest radiographs

被引:208
作者
Samei, E [1 ]
Flynn, MJ [1 ]
Eyler, WR [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Diagnost Radiol, Detroit, MI 48202 USA
关键词
diagnostic radiology; observer performance; lung neoplasms; diagnosis; lung; nodule; radiography; digital; receiver operating characteristic (ROC) curve; thorax;
D O I
10.1148/radiology.213.3.r99dc19727
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To assess the relative influence of quantum mottle and structured lung patterns (anatomic noise) on the detection of subtle lung nodules on chest radiographs. MATERIALS AND METHODS: Sixty 8 x 8-cm lung pattern images were extracted from digital chest radiographs in healthy individuals. Sixty quantum mottle images of the same size and quantum noise level were extracted from uniformly exposed digital radiographs. Simulated nodules with various peak contrast-diameter products (CD) that emulated subtle tissue-equivalent lung nodules were numerically superimposed at the center on three-fourths of the images. Printouts were independently viewed and scored by five experienced radiologists. The area under the receiver operating characteristic curve (A(z)) was estimated as a measure of the detectability of the nodules. RESULTS: At a fixed observer performance level (eg, A(z) = 0.8), much smaller and lower-contrast nodules were detected on quantum mottle images (1-mm diameter, CD = 0.01 mm), compared with those on anatomic images (4.5-mm diameter, CD = 0.20 mm). The findings generally agreed with the signal-to-noise ratio calculations based on statistical observer models. CONCLUSION: The detection of subtle lung nodules on chest radiographs is limited by anatomic noise.
引用
收藏
页码:727 / 734
页数:8
相关论文
共 37 条
[21]   LUNG-CANCER DETECTED DURING A SCREENING-PROGRAM USING 4-MONTH CHEST RADIOGRAPHS [J].
MUHM, JR ;
MILLER, WE ;
FONTANA, RS ;
SANDERSON, DR ;
UHLENHOPP, MA .
RADIOLOGY, 1983, 148 (03) :609-615
[22]  
NAKAJIMA N, 1994, FUJI COMPUTED RADIOG, V3
[23]   Influence of scatter reduction on lesion signal-to-noise ratio and lesion detection in digital chest radiography [J].
Neitzel, U ;
Pralow, T ;
Schaefer-Prokop, C ;
Prokop, M .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :337-347
[24]   INFLUENCE OF STRUCTURED NOISE ON DETECTION OF RADIOLOGIC ABNORMALITIES [J].
REVESZ, G ;
KUNDEL, HL ;
GRABER, MA .
INVESTIGATIVE RADIOLOGY, 1974, 9 (06) :479-486
[25]  
Rolland J. P. Y., 1990, FACTORS INFLUENCING
[27]   A SIMPLE-MODEL COMBINING QUANTUM NOISE AND ANATOMICAL VARIATION IN RADIOGRAPHS [J].
RUTTIMANN, UE ;
WEBBER, RL .
MEDICAL PHYSICS, 1984, 11 (01) :50-60
[28]   A method for measuring the presampled MTF of digital radiographic systems using an edge test device [J].
Samei, E ;
Flynn, MJ ;
Reimann, DA .
MEDICAL PHYSICS, 1998, 25 (01) :102-113
[29]   Comparison of observer performance for real and simulated nodules in chest radiography [J].
Samei, E ;
Flynn, MJ ;
Beute, GH ;
Peterson, E .
IMAGE PERCEPTION - MEDICAL IMAGING 1996, 1996, 2712 :60-70
[30]   Physical measures of image quality in photostimulable phosphor radiographic systems [J].
Samei, E ;
Flynn, MJ .
PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 :328-338