On the accuracy of techniques for obtaining the calibration coefficient NK of 192Ir HDR brachytherapy sources

被引:34
作者
Mainegra-Hing, Ernesto [1 ]
Rogers, D. W. O.
机构
[1] Natl Res Council Canada, Ionizing Radiat Stand, Ottawa, ON K1A 0R6, Canada
[2] Carleton Univ, Dept Phys, Ottawa, ON K1S 5B6, Canada
关键词
Ir-192; HDR; brachytherapy; calibration procedures;
D O I
10.1118/1.2239198
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The accuracy of interpolation or averaging procedures for obtaining the calibration coefficient N-K for Ir-192 high-dose-rate brachytherapy sources has been investigated using the EGSnrc Monte Carlo simulation system. It is shown that the widely used two-point averaging procedure of Goetsch et al. [Med. Phys. 18, 462 (1991)] has some conceptual problems. Most importantly, they recommended, as did the IAEA, averaging A(wall)N(K) values whereas one should average 1/N-K values. In practice this and other issues are shown to have little effect except for Goetsch et al.'s methods for determining A(wall) values. Their method of generalizing the A(wall) values measured in one geometry to other geometries is incorrect by up to 2%. However, these errors in A(wall) values cause systematic errors of only 0.3% in Ir-192 calibration coefficients. It is shown that A(wall) values need not be included in the averaging technique at all, thereby simplifying the technique considerably. It is demonstrated that as long as ion chambers with a flat response are used and/or very heavily filtered 250 kV (or higher) beams of x rays are used in the averaging, then almost all techniques can provide adequate accuracy. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:3340 / 3347
页数:8
相关论文
共 33 条
[1]   DESIGN OF A THIMBLE CHAMBER FOR FARMER DOSIMETER [J].
AIRD, EGA ;
FARMER, FT .
PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (02) :169-&
[2]  
Azario L, 1998, PHYS MEDICA, V14, P25
[3]   Spectra and air-kerma strength for encapsulated 192Ir sources [J].
Borg, J ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (11) :2441-2444
[4]   Monte Carlo study of correction factors for Spencer-Attix cavity theory at photon energies at or above 100 keV [J].
Borg, J ;
Kawrakow, I ;
Rogers, DWO ;
Seuntjens, JP .
MEDICAL PHYSICS, 2000, 27 (08) :1804-1813
[5]  
Borg J., 1998, PIRS629R NRC, DOI DOI 10.4224/5764262
[6]   CSnrc: Correlated sampling Monte Carlo calculations using EGSnrc [J].
Buckley, LA ;
Kawrakow, I ;
Rogers, DWO .
MEDICAL PHYSICS, 2004, 31 (12) :3425-3435
[7]   Results supporting calculated wall correction factors for cavity chambers [J].
Büermann, L ;
Kramer, HM ;
Csete, I .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (21) :3581-3594
[8]   CALIBRATION OF IR-192 HIGH-DOSE-RATE AFTERLOADING SYSTEMS [J].
GOETSCH, SJ ;
ATTIX, FH ;
PEARSON, DW ;
THOMADSEN, BR .
MEDICAL PHYSICS, 1991, 18 (03) :462-467
[9]  
IAEA, 2002, IAEA-TECDOC-1274, DOI [10.1016/s0167-8140(02)82497-5, DOI 10.1016/S0167-8140(02)82497-5]
[10]  
ICRU, 1984, 37 ICRU