Development of an assay for the simultaneous determination of sildenafil (Viagra) and its metabolite (UK-103,320) using automated sequential trace enrichment of dialysates and high-performance liquid chromatography

被引:102
作者
Cooper, JDH [1 ]
Muirhead, DC [1 ]
Taylor, JE [1 ]
Baker, PR [1 ]
机构
[1] PFIZER LTD,CENT RES,EARLY CLIN RES GRP,SANDWICH CT13 9NJ,KENT,ENGLAND
来源
JOURNAL OF CHROMATOGRAPHY B | 1997年 / 701卷 / 01期
关键词
trace enrichment; sildenafil; UK-103,320;
D O I
10.1016/S0378-4347(97)00339-3
中图分类号
Q5 [生物化学];
学科分类号
071010 [生物化学与分子生物学]; 081704 [应用化学];
摘要
The development of the ASTED (automated sequential trace enrichment of dialysates) system to prepare plasma samples prior to high-performance liquid chromatography (HPLC) of sildenafil and its demethylated metabolite (UK-103,320) is described. Investigations to elucidate potential pitfalls of the ASTED on-line sample preparation system prior to separation of the analytes by HPLC are presented. The procedure is shown to be selective for sildenafil and UK-103,320, and linear over the range 1.00-250 ng/ml. The intra-batch imprecision (C.V.) of the method at plasma analyte concentrations of 1.00, 5.00, 50.0 and 200 ng/ml was 11.2, 3.10, 1.50, and 1.30%, respectively, and the corresponding inter-batch imprecision was estimated to be 13.5, 7.09, 3.69, and 5.43%. At these plasma analyte concentrations the overall inaccuracy (% bias) of the procedure ranged from 3.6 to 8.4%. The method showed similar assay performance for the estimation of the metabolite, UK-103,320. The application of the assay to a pharmacokinetic investigation during a clinical study is presented. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:87 / 95
页数:9
相关论文
共 9 条
[1]
ONLINE COMBINATION OF DIALYSIS AND COLUMN-SWITCHING LIQUID-CHROMATOGRAPHY AS A FULLY AUTOMATED SAMPLE PREPARATION TECHNIQUE FOR BIOLOGICAL SAMPLES - DETERMINATION OF NITROFURAN RESIDUES IN EDIBLE PRODUCTS [J].
AERTS, MML ;
BEEK, WMJ ;
BRINKMAN, UAT .
JOURNAL OF CHROMATOGRAPHY, 1990, 500 :453-468
[2]
Boolell M, 1996, Int J Impot Res, V8, P47
[3]
AUTOMATED SEQUENTIAL TRACE ENRICHMENT OF DIALYSATES AND ROBOTICS - A TECHNIQUE FOR THE PREPARATION OF BIOLOGICAL SAMPLES PRIOR TO HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
COOPER, JDH ;
TURNELL, DC ;
GREEN, B ;
VERILLON, F .
JOURNAL OF CHROMATOGRAPHY, 1988, 456 (01) :53-69
[4]
COOPER JDH, 1994, LAB ROBOTICS AUTOMAT, V6, P15
[5]
AUTOMATIC PREPARATION OF HUMAN-SERUM SAMPLES FOR ANALYSIS OF THE DRUG ENOXIMONE AND ITS SULFOXIDE METABOLITE USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
COOPER, JDH ;
TURNELL, DC .
JOURNAL OF CHROMATOGRAPHY, 1986, 380 (01) :109-116
[6]
AUTOMATED SEQUENTIAL TRACE ENRICHMENT OF DIALYSATES COMBINED WITH HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND AUTOMATED HEART-CUTTING FOR THE DETERMINATION OF THE NUCLEOSIDE 1-(BETA-D-ARABINOFURANOSYL)-5-(1-PROPYNYL)URACIL AND ITS METABOLITE 5-PROPYNYLURACIL IN URINE [J].
COOPER, JDH ;
SHEUNG, CTCF ;
BUICK, AR .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 652 (01) :15-21
[7]
AN AUTOMATED-METHOD FOR THE ANALYSIS OF URINARY FREE CATECHOLAMINES USING ASTED AND HIGH-PRESSURE LIQUID-CHROMATOGRAPHY [J].
GREEN, B ;
COOPER, JDH ;
TURNELL, DC .
ANNALS OF CLINICAL BIOCHEMISTRY, 1989, 26 :361-367
[8]
ADSORPTION OF SERUM CALCIUM BY PLASTIC SAMPLE CUPS [J].
HALL, RA ;
WHITEHEAD, TP .
JOURNAL OF CLINICAL PATHOLOGY, 1970, 23 (04) :323-+
[9]
PREPARATIVE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF PEPTIDES ON A NEW REVERSED-PHASE PACKING MATERIAL, KROMASIL C-18 [J].
LARSSON, K ;
HERMANN, W ;
MOLLER, P ;
SANCHEZ, D .
JOURNAL OF CHROMATOGRAPHY, 1988, 450 (01) :71-80