Chronic evaluation of the Cleveland Clinic CorAide left ventricular assist system in calves

被引:16
作者
Fukamachi, K
Ochiai, Y
Doi, K
Massiello, AL
Medvedev, AL
Horvath, DJ
Gerhart, RL
Chen, JF
Krogulecki, AY
Takagaki, M
Howard, MW
Kopcak, MW
Golding, LAR
机构
[1] Cleveland Clin Fdn, Lerner Res Inst, Dept Biomed Engn, Cleveland, OH 44195 USA
[2] Cleveland Clin Fdn, Lerner Res Inst, Biol Resources Unit, Cleveland, OH 44195 USA
关键词
centrifugal pump; left ventricular assist device; anticoagulation; hemolysis;
D O I
10.1046/j.1525-1594.2002.06994.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Cleveland Clinic CorAide left ventricular assist system is based on a third-generation, implantable, centrifugal pump in which a rotating assembly is suspended fully. To evaluate chronic in vivo system performance and biocompatibility, the CorAide blood pump was implanted in 18 calves for either 1 month or 3 months. Hemodynamics were stable in all calves with a mean pump flow of 5.9 +/- 1.2 L/min and a mean systemic arterial pressure of 98 +/- 5 mm Hg. There were no incidences of bleeding, organ dysfunction, or mechanical failure in any of the 18 calves. Hemolysis occurred in only 1 calf due to outflow graft stenosis. Thrombus inside the pump, seen in 4 of the first 6 cases, was totally eliminated by a final redesign in the remaining cases, including the last 6 implants conducted without anticoagulation therapy. The CorAide blood pump demonstrated good biocompatibility and reliable, effective system performance.
引用
收藏
页码:529 / 533
页数:5
相关论文
共 12 条
[1]   Research and development of an implantable, axial-flow left ventricular assist device: The Jarvik 2000 heart [J].
Frazier, OH ;
Myers, TJ ;
Jarvik, RK ;
Westaby, S ;
Pigott, DW ;
Gregoric, ID ;
Khan, T ;
Tamez, DW ;
Conger, JL ;
Macris, MP .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S125-S132
[2]   HeartMate II left ventricular assist system: From concept to first clinical use [J].
Griffith, BP ;
Kormos, RL ;
Borovetz, HS ;
Litwak, K ;
Antaki, JF ;
Poirier, VL ;
Butler, KC .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S116-S120
[3]   The CorAide blood pump [J].
Horvath, DJ ;
Golding, LAR ;
Massiello, A ;
Medvedev, AL ;
Gerhart, RL ;
Ochiai, Y ;
Krogulecki, A ;
Fukamachi, K .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S191-S191
[4]   The HeartMate III: Design and in vivo studies of a Maglev centrifugal left ventricular assist device [J].
Loree, HM ;
Bourque, K ;
Gernes, DB ;
Richardson, JS ;
Poirier, VL ;
Barletta, N ;
Fleischli, A ;
Foiera, G ;
Gempp, TM ;
Schoeb, R ;
Litwak, KN ;
Akimoto, T ;
Kameneva, M ;
Watach, MJ ;
Litwak, P .
ARTIFICIAL ORGANS, 2001, 25 (05) :386-391
[5]   One hundred patients with the HeartMate left ventricular assist device: Evolving concepts and technology [J].
McCarthy, PM ;
Smedira, NO ;
Vargo, RL ;
Goormastic, M ;
Hobbs, RE ;
Starling, RC ;
Young, JB .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1998, 115 (04) :904-912
[6]   Clinical experience with the MicroMed DeBakey ventricular assist device [J].
Noon, GP ;
Morley, DL ;
Irwin, S ;
Abdelsayed, SV ;
Benkowski, RJ ;
Lynch, BE .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S133-S138
[7]   In vivo hemodynamic performance of the Cleveland Clinic CorAide blood pump in calves [J].
Ochiai, Y ;
Golding, LAR ;
Massiello, AL ;
Medvedev, AL ;
Gerhart, RL ;
Chen, JF ;
Takagaki, M ;
Fukamachi, K .
ANNALS OF THORACIC SURGERY, 2001, 72 (03) :747-752
[8]  
OCHIAI Y, IN PRESS ASAIO J
[9]   The totally implantable Novacor Left Ventricular Assist System [J].
Robbins, RC ;
Kown, MH ;
Portner, PM ;
Oyer, PE .
ANNALS OF THORACIC SURGERY, 2001, 71 (03) :S162-S165
[10]   The REMATCH trial: Rationale, design, and end points [J].
Rose, EA ;
Moskowitz, AJ ;
Packer, M ;
Sollano, JA ;
Williams, DL ;
Tierney, AR ;
Heitjan, DF ;
Meier, P ;
Ascheim, DD ;
Levitan, RG ;
Weinberg, AD ;
Stevenson, LW ;
Shapiro, PA ;
Lazar, RM ;
Watson, JT ;
Goldstein, DJ ;
Gelijns, AC .
ANNALS OF THORACIC SURGERY, 1999, 67 (03) :723-730