Scatter rejection in scanned multi-slit digital mammography

被引:12
作者
Åslund, M [1 ]
Cederström, B [1 ]
Lundqvist, M [1 ]
Danielsson, M [1 ]
机构
[1] Mamea Imaging AB, S-11428 Stockholm, Sweden
来源
MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2004年 / 5368卷
关键词
mammography digital; scatter; rejection; scanning; slit;
D O I
10.1117/12.535382
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Measurements and Monte Carlo simulations were used to investigate the scatter properties of a scanned multi-slit digital mammography system. Scatter to primary ratio (S/P) in the center of the image field was calculated for different thickness of breast equivalent material and different tube potentials. The simulated model also varied the angular acceptance, the number of slits and the distance between the slits of a dedicated scatter rejection device. In addition to the expected scatter from the breast equivalent material, scatter within the detector contributes to the S/P-ratio. The main part of the scatter is identified as coming from this process. Measured total S/P-ratios below 3% are reported for breast range 3 - 8 cm. The scatter-DQE is used as figure-of-merit for comparison to other imaging geometries and scatter rejection schemes.
引用
收藏
页码:478 / 487
页数:10
相关论文
共 9 条
[1]   Scatter/primary in mammography: Comprehensive results [J].
Boone, JM ;
Lindfors, KK ;
Cooper, VN ;
Seibert, JA .
MEDICAL PHYSICS, 2000, 27 (10) :2408-2416
[2]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[3]   The effect of the antiscatter grid on full-field digital mammography phantom images [J].
Chakraborty, DP .
JOURNAL OF DIGITAL IMAGING, 1999, 12 (01) :12-22
[4]   CALCULATION OF DOSE AND CONTRAST FOR 2 MAMMOGRAPHIC GRIDS [J].
DANCE, DR ;
PERSLIDEN, J ;
CARLSSON, GA .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :235-248
[5]   Scattered radiation in scanning slot mammography [J].
Jing, ZX ;
Huda, W ;
Walker, JK .
MEDICAL PHYSICS, 1998, 25 (07) :1111-1117
[6]   Scatter reduction in mammography with air gap [J].
Krol, A ;
Bassano, DA ;
Chamberlain, CC ;
Prasad, SC .
MEDICAL PHYSICS, 1996, 23 (07) :1263-1270
[7]   Mammography grid performance [J].
Rezentes, PS ;
de Almeida, A ;
Barnes, GT .
RADIOLOGY, 1999, 210 (01) :227-232
[8]   The value of scatter removal by a grid in full field digital mammography [J].
Veldkamp, WJH ;
Thijssen, MAO ;
Karssemeijer, N .
MEDICAL PHYSICS, 2003, 30 (07) :1712-1718
[9]   EXPERIMENTAL MEASUREMENTS OF THE SCATTER REDUCTION OBTAINED IN MAMMOGRAPHY WITH A SCANNING MULTIPLE SLIT ASSEMBLY [J].
YESTER, MV ;
BARNES, GT ;
KING, MA .
MEDICAL PHYSICS, 1981, 8 (02) :158-162