Scatter/primary in mammography: Comprehensive results

被引:138
作者
Boone, JM [1 ]
Lindfors, KK [1 ]
Cooper, VN [1 ]
Seibert, JA [1 ]
机构
[1] Univ Calif Davis, Dept Radiol, Sacramento, CA 95817 USA
关键词
scatter; mammography; breast imaging; Monte Carlo; computer simulation;
D O I
10.1118/1.1312812
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Monte Carlo procedures using the SIERRA code (validated in a companion article) were used to investigate the scatter properties in mammography. The scatter to primary ratio (SPR) was used for quantifying scatter levels as a function of beam spectrum, position in the field, air gap, breast thickness, tissue composition, and the area of the field of view (FOV). The geometry of slot scan mammography was also simulated, and SPR values were calculated as a function of slot width. The influence of large air gaps (to 30 cm) was also studied in the context of magnification mammography. X-ray energy and tissue composition from 100% adipose to 100% glandular demonstrated little effect on the SPR. Air gaps over a range from 0 to 30 mm showed only slight effects. The SPR increased with increased breast thickness and with larger fields of view. Measurements from 82 mammograms provided estimates of the range of compressed breast thickness (median: 5.2 cm, 95% range: 2.4 cm to 7.9 cm) and projected breast area onto the film (left craniocaudal view, median: 146 cm(2), 95% range: 58 cm(2) to 298 cm(2)). SPR values for semicircular breast shapes, Mo/Mo spectra, and a 15 mm air gap were parametrized as a function of breast thickness and (semicircular) breast diameter. With the coefficients a = - 2.35452817439093, b = 22.3960980055927, and c = 8.85064260299289, the equation SPR = [a + b x (diameter in cm) <^>(- 1.5) + c x (thickness in cm) <^>(- 0.5)] <^>- 1 produces SPR data from 2 to 8 cm and from 3 to 30 cm breast diameters with an average error of about 1%. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)02610-9].
引用
收藏
页码:2408 / 2416
页数:9
相关论文
共 16 条
[1]   AAPM TUTORIAL - CONTRAST AND SCATTER IN X-RAY-IMAGING [J].
BARNES, GT .
RADIOGRAPHICS, 1991, 11 (02) :307-323
[2]  
BARNES GT, 1993, MED PROG TECHNOL, V19, P7
[3]   INTENSITY OF SCATTERED RADIATION IN MAMMOGRAPHY [J].
BARNES, GT ;
BREZOVICH, IA .
RADIOLOGY, 1978, 126 (01) :243-247
[4]   Scatter/primary in mammography: Monte Carlo validation [J].
Boone, JM ;
Cooper, VN .
MEDICAL PHYSICS, 2000, 27 (08) :1818-1831
[5]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[6]   Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV [J].
Boone, JM ;
Seibert, JA .
MEDICAL PHYSICS, 1997, 24 (11) :1661-1670
[7]   CALCULATION OF DOSE AND CONTRAST FOR 2 MAMMOGRAPHIC GRIDS [J].
DANCE, DR ;
PERSLIDEN, J ;
CARLSSON, GA .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :235-248
[8]   THE COMPUTATION OF SCATTER IN MAMMOGRAPHY BY MONTE-CARLO METHODS [J].
DANCE, DR ;
DAY, GJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (03) :237-247
[9]   PERFORMANCE OF GLASS-FIBER ANTISCATTER DEVICES AT MAMMOGRAPHIC ENERGIES [J].
FAHRIG, R ;
MAINPRIZE, JG ;
ROBERT, N ;
ROGERS, A ;
YAFFE, MJ .
MEDICAL PHYSICS, 1994, 21 (08) :1277-1282
[10]   SCATTER PRIMARY RATIOS FOR X-RAY-SPECTRA MODIFIED TO ENHANCE IODINE CONTRAST IN SCREEN FILM MAMMOGRAPHY [J].
FRITZ, SL ;
CHANG, CHJ ;
LIVINGSTON, WH .
MEDICAL PHYSICS, 1983, 10 (06) :866-870