Glow-discharge treated piezoelectric quartz crystals as immunosensors for HSA detection

被引:31
作者
Saber, R
Mutlu, S
Piskin, E [1 ]
机构
[1] Hacettepe Univ, Dept Chem Engn, TR-06532 Ankara, Turkey
[2] Hacettepe Univ, Bioengn Div, TR-06532 Ankara, Turkey
关键词
immunosensor; piezoelectric quartz crystal; glow-discharge; antibody immobilization; human serum albumin (HSA);
D O I
10.1016/S0956-5663(02)00058-1
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The aim of this study is to develop an immunoaffinity sensor based on piezoelectric crystals for human serum albumin (HSA) detection in aqueous media. Quartz crystals were treated with ethylene diamine (EDA) plasma in a glow-discharge apparatus in order to substitute amino groups on their surfaces. Then anti-HSA antibodies were immobilized via these amino groups by using glutaraldehyde (GA) as cross linker. Immobilization of the antibody on the quartz crystal was examined for different pH, antibody concentration and treatment time. The optimum conditions for anti-HSA immobilization were evaluated by the measurements of the activity of the surface against HSA. The optimum values of pH, antibody concentration and treatment time were found 6.2, 0.15 mg/ml and 2 h, respectively. For detection of HSA into the solution, two methods were used. In the first (dip and dry) method, the frequency shifts were measured in air after the 1 h interaction of the anti-HSA immobilized crystals with HSA solution. In the other (direct) method, the frequency shifts were followed continuously for 60 min. while the probe was immersed in the HSA solution. An increase for the frequency shifts was observed with increasing of HSA concentration of 16-128 mug/ml. Both the immobilization and antibody-interaction conditions were found important on the extend of these specific interaction. The relations between the HSA concentrations and frequency shifts were exponential in both methods. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:727 / 734
页数:8
相关论文
共 37 条
[1]   A quartz crystal microbalance assay for detection of antibodies against the recombinant African swine fever virus attachment protein p12 in swine serum [J].
Abad, JM ;
Pariente, F ;
Hernández, L ;
Lorenzo, E .
ANALYTICA CHIMICA ACTA, 1998, 368 (03) :183-189
[2]   Evaluation of antibody immobilization methods for piezoelectric biosensor application [J].
Babacan, S ;
Pivarnik, P ;
Letcher, S ;
Rand, AG .
BIOSENSORS & BIOELECTRONICS, 2000, 15 (11-12) :615-621
[3]   STUDY OF PERFORMANCE AND CHEMICAL CHARACTERISTICS OF COMPOSITE REVERSE-OSMOSIS MEMBRANES PREPARED BY PLASMA POLYMERIZATION OF ALLYLAMINE [J].
BELL, AT ;
WYDEVEN, T ;
JOHNSON, CC .
JOURNAL OF APPLIED POLYMER SCIENCE, 1975, 19 (07) :1911-1930
[4]   SENSORS BASED ON PIEZOELECTRIC RESONATORS [J].
BENES, E ;
GROSCHL, M ;
BURGER, W ;
SCHMID, M .
SENSORS AND ACTUATORS A-PHYSICAL, 1995, 48 (01) :1-21
[5]  
Chen HM, 1999, J CHIN INST CHEM ENG, V30, P413
[6]   Piezoelectric cryptand-coated quartz crystal liquid chromatographic detector for cations anions and polar organic molecules in solutions [J].
Chiou, CS ;
Shih, JS .
ANALYTICA CHIMICA ACTA, 1999, 392 (2-3) :125-133
[7]   On-line immunoanalysis for environmental pollutants:: from batch assays to automated sensors [J].
González-Martínez, MA ;
Puchades, R ;
Maquieira, A .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1999, 18 (03) :204-218
[8]  
HERMANSON GT, 1996, BIOCONJUGATE TECHNIQ, P460
[9]   SPECTROSCOPIC ELUCIDATION OF CHEMICAL-STRUCTURE OF PLASMA-POLYMERIZED PYRIDINE [J].
HOZUMI, K ;
KITAMURA, K ;
HASHIMOTO, H ;
HAMAOKA, T ;
FUJISAWA, H ;
ISHIZAWA, T .
JOURNAL OF APPLIED POLYMER SCIENCE, 1983, 28 (05) :1651-1665
[10]  
KONIG B, 1993, ANAL LETT, V26, P2313