Symptomatic pericardial disease associated with patch electrodes of the automatic implantable cardioverter defibrillator: An underestimated complication?

被引:19
作者
Chevalier, P [1 ]
Moncada, E [1 ]
Canu, G [1 ]
Claudel, JP [1 ]
Bellon, C [1 ]
Kirkorian, G [1 ]
Touboul, P [1 ]
机构
[1] HOP CARDIOVASC & PNEUMOL LOUIS PRADEL,F-69394 LYON 03,FRANCE
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1996年 / 19卷 / 12期
关键词
implantable defibrillator; constrictive pericarditis;
D O I
10.1111/j.1540-8159.1996.tb03292.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Constrictive pericarditis can be associated with ICD patch electrodes. During a mean follow-up of 24 months, in a population of 35 patients who received ICDs with a patch electrodes configuration, ne identified three patients with clinical and hemodynamic signs compatible with this event. patient 1, a 35-year-old male, underwent implantation of an ICD because of a primary electrical disease complicated by cardiac arrest. Fourteen months later he complained of exertional dyspnea without any signs of heart failure. Right heart catheterization show ed high filling pressures and diastolic dip and plateau in pressure curves. Thoracotomy and pericardial exploration rr ere performed. Three months after removal of the patches and insertion of an endocardial lead system, the patient had normal respiration. patients 2 and 3, who suffered from coronary heart disease without heart failure, exhibited a hemodynamic profile suggestive of constrictive pericarditis: in one patient, 10 months after ICD implantation, associated with right heart failure; and in the other 18 months after TCD implantation with left heart failure. Patch electrodes were removed in these two patients and replaced by endocardial lead electrodes with subsequent clinical improvement. I is concluded that constrictive pericarditis related to epicardial patch is not an uncommon occurrence during ICD therapy and should be considered in patients rr ho show clinical signs of cardiac decompensation.
引用
收藏
页码:2150 / 2152
页数:3
相关论文
共 6 条
[1]   CONSTRICTIVE PERICARDITIS ASSOCIATED WITH PATCH ELECTRODES OF THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR [J].
ALMASSI, GH ;
CHAPMAN, PD ;
TROUP, PJ ;
WETHERBEE, JN ;
OLINGER, GN .
CHEST, 1987, 92 (02) :369-371
[2]   EFFECT OF AICD PATCH ELECTRODES ON THE DIASTOLIC PRESSURE VOLUME CURVE IN PIGS [J].
AUTERI, JS ;
JEEVANANDAM, V ;
BIELEFELD, MR ;
SANCHEZ, JA ;
SPOTNITZ, HM .
ANNALS OF THORACIC SURGERY, 1991, 52 (05) :1052-1057
[3]   IMPLANTABLE CARDIOVERTER DEFIBRILLATOR INFECTION CAUSING CONSTRICTIVE PERICARDITIS [J].
KASSANOFF, AH ;
LEVIN, CB ;
WYNDHAM, CRC ;
MILLS, LJ .
CHEST, 1992, 102 (03) :960-963
[4]   CONSTRICTIVE PERICARDIAL DISEASE CAUSED BY EPICARDIAL IMPLANTABLE CARDIAC DEFIBRILLATOR PATCHES - TREATMENT BY PERICARDIAL STRIPPING AND NONTHORACOTOMY LEAD SYSTEM IMPLANTATION [J].
LURIE, K ;
SHULTZ, J ;
REMOLE, S ;
ASSO, A ;
PINEDA, E ;
BENDITT, D ;
FETTER, J ;
LAXSON, D .
AMERICAN HEART JOURNAL, 1994, 128 (03) :623-625
[5]   PATHOLOGICAL FINDINGS RELATED TO THE LEAD SYSTEM AND REPEATED DEFIBRILLATIONS IN PATIENTS WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR [J].
SINGER, I ;
HUTCHINS, GM ;
MIROWSKI, M ;
MOWER, MM ;
VELTRI, EP ;
GUARNIERI, T ;
GRIFFITH, LSC ;
WATKINS, L ;
JUANTEGUY, J ;
FISHER, S ;
REID, PR ;
WEISFELDT, ML .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (02) :382-388
[6]  
WALSH TJ, 1982, J THORAC CARDIOVASC, V8, P126