Residual glutaraldehyde levels in fiberoptic endoscopes: measurement and implications for patient toxicity

被引:14
作者
Farina, A
Fievet, MH
Plassart, F
Menet, MC
Thuillier, A
机构
[1] Grp Hosp Pitie Salpetriere, F-75013 Paris, France
[2] Fac Pharm, Chim Analyt Lab, F-75005 Paris, France
关键词
glutaraldehyde; toxicity; fiberoptic endoscopes; disinfection;
D O I
10.1016/S0195-6701(99)90425-5
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Most gastroenterology societies recommend glutaraldehyde for fiberoptic endoscope disinfection. However, glutaraldehyde toxicity has been suspected in patients examined with endoscopes disinfected with this compound. The aim of our study was to determine the residual levels of glutaraldehyde in fiberoptic endoscopes after either manual or automatic disinfection and to evaluate the extent of toxicity Furthermore, the procedures for disinfection currently performed by the department a ere compared with the new French guidelines. We used both manual and automatic disinfection procedures and flushed sterile distilled water through the lumens of endoscopes before use. Residual glutaraldehyde levels were determined using liquid chromatography coupled to spectrophotometric detection. In a total of 92 measurements it was found that residual glutaraldehyde levels were higher and more variable after manual disinfection (< 0.2-159.5 mg/L) than after automatic disinfection (< 0.2-6.3 mg/L). We conclude that local procedures for disinfection need to be improved to conform to the new French guidelines. Since thresholds for the toxic dose of glutaraldehyde and international norms for levels of residual glutaraldehyde in equipment have not been defined, additional studies combining accurate measurements in fiberoptic endoscopes and clinical observations of endoscopy patients will be required to draw more definitive conclusions. (C) 1999 The Hospital Infection Society.
引用
收藏
页码:293 / 297
页数:5
相关论文
共 16 条
[1]   GLUTARALDEHYDE-INDUCED PROCTITIS AFTER ENDORECTAL ULTRASOUND EXAMINATION - A HIGHER RISK OF INCIDENCE THAN EXPECTED [J].
BURTIN, P ;
RUGET, O ;
PETIT, R ;
BOYER, J .
GASTROINTESTINAL ENDOSCOPY, 1993, 39 (06) :859-860
[2]   Acute colitis following colonoscopy [J].
Caprilli, R ;
Viscido, A ;
Frieri, G ;
Latella, G .
ENDOSCOPY, 1998, 30 (04) :428-431
[3]   OCCUPATIONAL ASTHMA IN RADIOGRAPHERS [J].
CULLINAN, P ;
HAYES, J ;
CANNON, J ;
MADAN, I ;
HEAP, D ;
TAYLOR, AN .
LANCET, 1992, 340 (8833) :1477-1477
[4]  
deLedinghen V, 1996, GASTROEN CLIN BIOL, V20, P215
[5]  
DOCE P, 1995, AM J INFECT CONTROL, V23, P34
[6]   INVESTIGATION OF AN OUTBREAK OF BLOODY DIARRHEA - ASSOCIATION WITH ENDOSCOPIC CLEANING SOLUTION AND DEMONSTRATION OF LESIONS IN AN ANIMAL-MODEL [J].
DURANTE, L ;
ZULTY, JC ;
ISRAEL, E ;
POWERS, PJ ;
RUSSELL, RG ;
QIZILBASH, AH ;
MORRIS, JG .
AMERICAN JOURNAL OF MEDICINE, 1992, 92 (05) :476-480
[7]   CIDEX-INDUCED SYNOVITIS [J].
HARNER, CD ;
FU, FH ;
MASON, GC ;
WISSINGER, HA ;
RABIN, BS .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1989, 17 (01) :96-102
[8]  
*HLTH SAF EX GUID, 1991, EH4091 HLTH SAF EX G
[9]  
LEMOAN G, 1983, ANN FALSIF EXP CHIM, V76, P463
[10]   Fast specific separation and sensitive quantification of bactericidal and sporicidal aldehydes by high-performance liquid chromatography: Example of glutaraldehyde determination [J].
Menet, MC ;
Gueylard, D ;
Fievet, MH ;
Thuillier, A .
JOURNAL OF CHROMATOGRAPHY B, 1997, 692 (01) :79-86