Dose perturbation effects in prostate seed implant brachytherapy with I-125

被引:22
作者
Mobit, P [1 ]
Badragan, I
机构
[1] Univ Calgary, Fac Med, Dept Oncol, Calgary, AB T2N 1N4, Canada
[2] Tom Baker Canc Clin, Dept Med Phys, Calgary, AB T2N 4N2, Canada
[3] Harper Univ Hosp, Ann Barbara Karmanos Canc Inst, Gershenson Radiat Oncol Ctr, Detroit, MI 48201 USA
[4] Wayne State Univ, Detroit, MI 48201 USA
关键词
D O I
10.1088/0031-9155/49/14/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
EGSnrc Monte Carlo simulation was used to investigate dose perturbation effects in prostate seed implant brachytherapy using I-125 radioactive seeds used in implant brachytherapy. Dose perturbation effects resulting from the seed mutual attenuation in a prostate seed implant consisting of 27 seeds were investigated. The results showed that for I-125 seeds implanted into the prostate at 1.00 cm, 0.75 cm and 0.50 cm apart (uniform spacing), the dose perturbation effects are up to 10%. The volume of the target occupied by the 10% dose difference between the full Monte Carlo simulation and the single seed superposition model decreases with increasing seed spacing. Despite the differences between the Monte Carlo simulation and the simple superposition, there was no significant change in the dose volume histogram for 1 cm and 0.75 cm seed spacing. However, there was a significant change in the dose volume histogram when the seed spacing was 0.5 cm. An analysis of the external volume index (El), coverage index (CI) and homogeneity index (HI) also showed that there is no difference in these indexes for the 1.00 cm and 0.75 cm seed spacing between the simple superposition model and the full Monte Carlo simulation. Compared to the full Monte Carlo simulations, the simple superposition model overestimated El, CI and HI by 7%, 5% and 4% respectively for the 0.50 cm seed spacing.
引用
收藏
页码:3171 / 3178
页数:8
相关论文
共 6 条
[1]
THE ACCURACY OF SINGLE-SEED DOSE SUPERPOSITION FOR I125 IMPLANTS [J].
BURNS, GS ;
RAESIDE, DE .
MEDICAL PHYSICS, 1989, 16 (04) :627-631
[2]
Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version [J].
Kawrakow, I .
MEDICAL PHYSICS, 2000, 27 (03) :485-498
[3]
INTERSEED EFFECTS ON DOSE FOR I-125 BRACHYTHERAPY IMPLANTS [J].
MEIGOONI, AS ;
MELI, JA ;
NATH, R .
MEDICAL PHYSICS, 1992, 19 (02) :385-390
[4]
An evaluation of the AAPM-TG43 dosimetry protocol for I-125 brachytherapy seed [J].
Mobit, P ;
Badragan, I .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (14) :3161-3170
[5]
Correlation of medical dosimetry quality indicators to the local tumor control in patients with prostate cancer treated with iodine-125 interstitial implants [J].
Nath, R ;
Roberts, K ;
Ng, M ;
Peschel, R ;
Chen, Z .
MEDICAL PHYSICS, 1998, 25 (12) :2293-2307
[6]
DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43 [J].
NATH, R ;
ANDERSON, LL ;
LUXTON, G ;
WEAVER, KA ;
WILLIAMSON, JF ;
MEIGOONI, AS .