Time Lag of Glucose From Intravascular to Interstitial Compartment in Humans

被引:116
作者
Basu, Ananda [1 ]
Dube, Simmi [1 ]
Slama, Michael [1 ]
Errazuriz, Isabel [1 ]
Amezcua, Jose Carlos [1 ]
Kudva, Yogish C. [1 ]
Peyser, Thomas [2 ]
Carter, Rickey E. [3 ]
Cobelli, Claudio [4 ]
Basu, Rita [1 ]
机构
[1] Mayo Clin, Endocrine Res Unit, Rochester, MN 55905 USA
[2] Dexcom Inc, San Diego, CA USA
[3] Mayo Coll Med, Dept Hlth Sci Res, Rochester, MN USA
[4] Univ Padua, Dept Informat Engn, Padua, Italy
基金
美国国家卫生研究院;
关键词
SUBCUTANEOUS ADIPOSE-TISSUE; TYPE-1; DIABETIC-PATIENTS; HEALTHY-VOLUNTEERS; ARTIFICIAL PANCREAS; INSULIN ACTION; MICRODIALYSIS; BLOOD; PATTERN; SENSOR; SYSTEM;
D O I
10.2337/db13-1132
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The accuracy of continuous interstitial fluid (ISF) glucose sensing is an essential component of current and emerging open- and closed-loop systems for type 1 diabetes. An important determinant of sensor accuracy is the physiological time lag of glucose transport from the vascular to the interstitial space. We performed the first direct measurement of this phenomenon to our knowledge in eight healthy subjects under an overnight fasted condition. Microdialysis catheters were inserted into the abdominal subcutaneous space. After intravenous bolus administrations of glucose tracers, timed samples of plasma and ISF were collected sequentially and analyzed for tracer enrichments. After accounting for catheter dead space and assay noise, the mean time lag of tracer appearance in the interstitial space was 5.3-6.2 min. We conclude that in the overnight fasted state in healthy adults, the physiological delay of glucose transport from the vascular to the interstitial space is 5-6 min. Physiological delay between blood glucose and ISF glucose, therefore, should not be an obstacle to sensor accuracy in overnight or fasting-state closed-loop systems of insulin delivery or open-loop therapy assessment for type 1 diabetes.
引用
收藏
页码:4083 / 4087
页数:5
相关论文
共 18 条
[1]   Effects of Type 2 Diabetes on Insulin Secretion, Insulin Action, Glucose Effectiveness, and Postprandial Glucose Metabolism [J].
Basu, Ananda ;
Man, Chiara Dalla ;
Basu, Rita ;
Toffolo, Gianna ;
Cobelli, Claudio ;
Rizza, Robert A. .
DIABETES CARE, 2009, 32 (05) :866-872
[2]   DETERMINATION OF THE C-13-LABELING PATTERN OF GLUCOSE BY GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
BEYLOT, M ;
PREVIS, SF ;
DAVID, F ;
BRUNENGRABER, H .
ANALYTICAL BIOCHEMISTRY, 1993, 212 (02) :526-531
[3]   Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor [J].
Boyne, MS ;
Silver, DM ;
Kaplan, J ;
Saudek, CD .
DIABETES, 2003, 52 (11) :2790-2794
[4]   Artificial Pancreas Past, Present, Future [J].
Cobelli, Claudio ;
Renard, Eric ;
Kovatchev, Boris .
DIABETES, 2011, 60 (11) :2672-2682
[5]  
Davey Raymond J, 2010, J Diabetes Sci Technol, V4, P1393
[6]   Correspondence of continuous interstitial glucose measurement against arterialised and capillary glucose following an oral glucose tolerance test in healthy volunteers [J].
Dye, Louise ;
Mansfield, Michael ;
Lasikiewicz, Nicola ;
Mahawish, Lena ;
Schnell, Rainer ;
Talbot, Duncan ;
Chauhan, Hitesh ;
Croden, Fiona ;
Lawton, Clare .
BRITISH JOURNAL OF NUTRITION, 2010, 103 (01) :134-140
[7]   Diurnal Pattern of Insulin Action in Type 1 Diabetes Implications for a Closed-Loop System [J].
Hinshaw, Ling ;
Dalla Man, Chiara ;
Nandy, Debashis K. ;
Saad, Ahmed ;
Bharucha, Adil E. ;
Levine, James A. ;
Rizza, Robert A. ;
Basu, Rita ;
Carter, Rickey E. ;
Cobelli, Claudio ;
Kudva, Yogish C. ;
Basu, Ananda .
DIABETES, 2013, 62 (07) :2223-2229
[8]   Effects of glucose and insulin levels on adipose tissue glucose measurement by microdialysis probes retained for three weeks in Type 1 diabetic patients [J].
Hullegie, LM ;
Lutgers, HL ;
Dullaart, RPF ;
Sluiter, WJ ;
Wientjes, KJ ;
Schoonen, AJM ;
Hoogenberg, K .
NETHERLANDS JOURNAL OF MEDICINE, 2000, 57 (01) :13-19
[9]   Subcutaneous a continuous glucose monitoring - Feasibility of a new microdialysis-based glucose sensor system [J].
Jungheim, K ;
Wientjes, KJ ;
Heinemann, L ;
Lodwig, V ;
Koschinsky, T ;
Koschinsky, T ;
Schoonen, AJ .
DIABETES CARE, 2001, 24 (09) :1696-1697
[10]  
Keenan D Barry, 2009, J Diabetes Sci Technol, V3, P1207