Correspondence of continuous interstitial glucose measurement against arterialised and capillary glucose following an oral glucose tolerance test in healthy volunteers

被引:20
作者
Dye, Louise [1 ]
Mansfield, Michael [2 ]
Lasikiewicz, Nicola [1 ]
Mahawish, Lena [2 ]
Schnell, Rainer [3 ]
Talbot, Duncan [4 ]
Chauhan, Hitesh [4 ]
Croden, Fiona [1 ]
Lawton, Clare [1 ]
机构
[1] Univ Leeds, Inst Psychol Sci, Human Appetite Res Unit, Leeds LS2 9JT, W Yorkshire, England
[2] St James Univ Hosp, Ctr Diabet, Leeds LS9 7TF, W Yorkshire, England
[3] Univ Duisburg Essen, D-47057 Duisburg, Germany
[4] Unilever R&D, Sharnbrook MK44 1LQ, Beds, England
关键词
Continuous glucose monitoring; Arterialised venous glucose; Capillary glucose; Validation; HOMEOSTASIS MODEL ASSESSMENT; DIABETIC-PATIENTS; PLASMA-GLUCOSE; GLYCEMIC INDEX; MICRODIALYSIS TECHNIQUE; QUESTIONNAIRE; ACCURACY; INSULIN;
D O I
10.1017/S0007114509991504
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
The aim of the present study was to validate the Glucoday(R) continuous interstitial ambulatory glucose-monitoring device (AGD) against plasma glucose measured from arterialised venous (AV) and glucose from capillary whole blood (finger prick, FP) in non-diabetic subjects in response to an oral glucose tolerance test. Fifteen healthy overweight men (age 30-49 years, BMI 26-31 kg/m(2)) participated. Glucose levels were measured before. during and after consumption Of an oral 75 g glucose load using twelve FP samples and forty-four 1 ml AV blood samples during 180 min. Interstitial glucose was measured via the AGD. Three venous samples for fasting insulin were taken to estimate insulin resistance. Profiles of AGD, AV and FP glucose were generated for each participant. Glucose values for each minute of the measurement period were interpolated using a locally weighted scatterplot smoother. Data were compared using Bland-Altman plots that showed good correspondence between all pairs of measurements. Concordance between the three methods was 0.8771 (Kendall's W, it 15 P<0.001). Concordance was greater between AV and FP (W = 0.9696) than AGD and AV (W = 0.8770) or AGD and FP (W = 0,8764). Analysis of time to peak glucose indicated that AGD measures lagged approximately 15 min behind FP and AV measures. Percent body fat was significantly correlated with time to peak glucose levels for each measure, while BMI and estimated insulin resistance (homeostatic model assessment, HOMA) were not. In conclusion, AGD shows good correspondence with FP and AV glucose measures in response to a glucose load with a 15 min time lag. Taking this into account, AGD has potential application in nutrition and behaviour studies.
引用
收藏
页码:134 / 140
页数:7
相关论文
共 29 条
[1]  
[Anonymous], 1999, Definition, diagnosis, and classification of diabetes mellitus and its complications: report of a WHO consultation
[2]  
[Anonymous], 1990, APPL NONPARAMETRIC R, DOI DOI 10.1017/CCOL0521382483
[3]   COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Glycaemic index methodology [J].
Brouns, F ;
Bjorck, I ;
Frayn, KN ;
Gibbs, AL ;
Lang, V ;
Slama, G ;
Wolever, TMS .
NUTRITION RESEARCH REVIEWS, 2005, 18 (01) :145-171
[6]   The beneficial effect of a diet with low glycaemic index on 24h glucose profiles in healthy young people as assessed by continuous glucose monitoring [J].
Brynes, AE ;
Adamson, J ;
Dornhorst, A ;
Frost, GS .
BRITISH JOURNAL OF NUTRITION, 2005, 93 (02) :179-182
[7]   Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis [J].
Clarke, WL ;
Anderson, S ;
Farhy, L ;
Breton, M ;
Gonder-Frederick, L ;
Cox, D ;
Kovatchev, B .
DIABETES CARE, 2005, 28 (10) :2412-2417
[8]  
Cleveland W.S., 1992, Statistical Models in S, P309, DOI DOI 10.1201/9780203738535-8
[9]  
JARRETT RJ, 1982, DIABETOLOGIA, V22, P79, DOI 10.1007/BF00254833
[10]   Evaluating the accuracy of continuous glucose-monitoring sensors [J].
Kovatchev, BP ;
Gonder-Frederick, LA ;
Cox, DJ ;
Clarke, WL .
DIABETES CARE, 2004, 27 (08) :1922-1928