Classification and Potential Mechanisms of Intravascular Ultrasound Patterns of Stent Fracture

被引:84
作者
Doi, Hiroshi [1 ]
Maehara, Akiko
Mintz, Gary S.
Tsujita, Kenichi
Kubo, Takashi
Castellanos, Celia
Liu, Jian
Yang, Junqing
Oviedo, Carlos
Aoki, Jiro
Franklin-Bond, Theresa
Dasgupta, Neil
Lansky, Alexandra J.
Dangas, George D.
Stone, Gregg W.
Moses, Jeffrey W.
Mehran, Roxana
Leon, Martin B.
机构
[1] Cardiovasc Res Fdn, New York, NY USA
关键词
SIROLIMUS-ELUTING STENT; CORONARY-ARTERY LESIONS; CLINICAL CHARACTERISTICS; IMPLANTATION; ANGIOPLASTY; ANEURYSM; APPOSITION; THROMBOSIS; FREQUENCY; OUTCOMES;
D O I
10.1016/j.amjcard.2008.11.051
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We sought to examine the intravascular ultrasound (IVUS) findings of stent fracture. Stent fracture has been implicated as a cause of drug-eluting stent failure. IVUS is more likely to identify mechanisms of stent failure-including stent fracture-than angiography. Twenty stent fractures diagnosed by IVUS in 17 patients were evaluated. Eighteen stent fractures (90%) occurred in sirolimus-eluting Cypher stents, and 2 stent fractures (10%) occurred in bare metal stents, but none occurred in paclitaxel-eluting Taxus stents. Half of the stent fractures presented <= 1 year after implantation, and 1/2 presented > 1 year after implantation. IVUS analysis showed that 9 stent fractures were complete (45%) and 11 were partial (55%); 10 (50%) were adjacent to stent metal overlap; and 5 occurred in a coronary aneurysm accompanied by malapposition (all Cypher stents) despite the absence of an aneurysm at index stenting. Compared with 60 matched control segments in patients without stent fracture, but with similar clinical events, the stent fracture group had longer stent segments (45.2 +/- 23.0 vs 28.5 +/- 14.9 mm, p = 0.003). Comparing stent fractures associated with an aneurysm (n = 5) with those that did not occur in association with an aneurysm (n = 15) showed that complete stent fracture was more frequent (100% vs 27%, p = 0.008), and all presented > 1 year after index stenting (vs 33%, p = 0.03). In conclusion, IVUS is helpful to identify stent fracture as a cause of stent failure and to understand possible mechanisms of stent fracture such as aneurysm formation. (C) 2009 Elsevier Inc. All rights reserved. (Am J Cardiol 2009;103:818-823)
引用
收藏
页码:818 / 823
页数:6
相关论文
共 29 条
[1]   Coronary Artery Aneurysms After Drug-Eluting Stent Implantation [J].
Aoki, Jiro ;
Kirtane, Ajay ;
Leon, Martin B. ;
Dangas, George .
JACC-CARDIOVASCULAR INTERVENTIONS, 2008, 1 (01) :14-21
[2]   Incidence and clinical impact of coronary stent fracture after sirolimus-eluting stent implantation [J].
Aoki, Jiro ;
Nakazawa, Gaku ;
Tanabe, Kengo ;
Hoye, Angela ;
Yamamoto, Hirosada ;
Nakayama, Tomohiro ;
Onuma, Yoshinobu ;
Higashikuni, Yasutomi ;
Otsuki, Syuji ;
Yagishita, Atsuhiko ;
Yachi, Sen ;
Nakajima, Hiroyoshi ;
Hara, Kazuhiro .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (03) :380-386
[3]  
Aziz Shahid, 2007, J Invasive Cardiol, V19, pE96
[4]   ACUTE COMPLICATIONS OF EXCIMER-LASER CORONARY ANGIOPLASTY - A DETAILED ANALYSIS OF MULTICENTER RESULTS [J].
BAUMBACH, A ;
BITTL, JA ;
FLECK, E ;
GESCHWIND, HJ ;
SANBORN, TA ;
TCHENG, JE ;
KARSCH, KR ;
COHEN, E ;
WARREN, S ;
MORRIS, D ;
SEGAL, J ;
GREEN, R ;
SIEGEL, R ;
BRINKER, J ;
KUCINSKI, C ;
BLONDER, R ;
WALTER, P ;
OVERLIE, P ;
SWAYE, P ;
VIGNOLA, P ;
ABELA, G ;
SANBORN, T ;
TOMASSO, C ;
KLEINER, J ;
MOOTHART, R ;
KEENE, F ;
BERGIN, P ;
WATSON, L ;
YARDLEY, D ;
ISNER, J ;
ROSENFIELD, K ;
HARRIS, B ;
POWER, J ;
TALBERT, CR ;
CANEDO, M ;
CHOKSHI, S ;
CHAPMAN, J ;
BUCHBINDER, M ;
TOBIS, J ;
GINSBURG, R ;
WINNIFORD, M ;
ZIMMER, S ;
WILSON, R ;
ELLIS, S ;
VANDERHEUVEL ;
REIFART, N ;
OZBEK, C ;
HOPP, HW ;
MARESTA, A ;
ESCOJIDO, H .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (06) :1305-1313
[5]   RELATION OF DEEP ARTERIAL RESECTION AND CORONARY-ARTERY ANEURYSMS AFTER DIRECTIONAL CORONARY ATHERECTOMY [J].
BELL, MR ;
GARRATT, KN ;
BRESNAHAN, JF ;
EDWARDS, WD ;
HOLMES, DR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (07) :1474-1481
[6]   The incidence and clinical impact of stent strut fractures developed after drug-eluting stent implantation [J].
Chung, Wook-Sung ;
Park, Chul-Soo ;
Seung, Ki-Bae ;
Kim, Pum-Joon ;
Lee, Jong-Min ;
Koo, Bon-Kwon ;
Jang, Yang-Soo ;
Yang, Joo-Young ;
Yoon, Jung-Han ;
Kim, Doo-Il ;
Yoon, Young-Won ;
Park, Jong-Seon ;
Cho, Yoon-Haeng ;
Park, Seung-Jung .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2008, 125 (03) :325-331
[7]   Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions [J].
Colombo, A ;
Drzewiecki, J ;
Banning, A ;
Grube, E ;
Hauptmann, K ;
Silber, S ;
Dudek, D ;
Fort, S ;
Schiele, F ;
Zmudka, K ;
Guagliumi, G ;
Russell, ME .
CIRCULATION, 2003, 108 (07) :788-794
[8]  
Condado JA, 1997, CIRCULATION, V96, P727
[9]   Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation [J].
Cook, Stephane ;
Wenaweser, Peter ;
Togni, Mario ;
Billinger, Michael ;
Morger, Cyrill ;
Seiler, Christian ;
Vogel, Rolf ;
Hess, Otto ;
Meier, Bernhard ;
Windecker, Stephan .
CIRCULATION, 2007, 115 (18) :2426-2434
[10]   Comparable clinical outcomes with paclitaxel- and sirolimus-eluting stents in unrestricted contemporary practice [J].
Cosgrave, John ;
Melzi, Gloria ;
Corbett, Simon ;
Biondi-Zoccai, Giuseppe G. L. ;
Agostoni, Pierfrancesco ;
Babic, Rade ;
Airoldi, Flavio ;
Chieffio, Alaide ;
Sangiorgi, Giuseppe M. ;
Montorfano, Matteo ;
Michev, Iassen ;
Carlino, Mauro ;
Colombo, Antonio .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (24) :2320-2328