Combined stent implantation and embolization with liquid 2-polyhydroxyethyl methacrylate for treatment of experimental canine wide-necked aneurysms

被引:15
作者
Klisch, J
Schellhammer, F
Zitt, J
Scheufler, KM
Pagenstecher, A
Nagursky, H
Schumacher, M
机构
[1] Univ Freiburg, Neuroradiol Sect, D-79106 Freiburg, Germany
[2] Univ Cologne, Dept Radiol, Cologne, Germany
[3] Univ Freiburg, Dept Neurosurg, Freiburg, Germany
[4] Univ Freiburg, Dept Neuropathol, Freiburg, Germany
[5] Univ Freiburg, Dept Oral & Maxillofacial Surg, Freiburg, Germany
关键词
aneurysm; stent; embolization; interventional neuroradiology; polyHEMA; hydrogel; animal study;
D O I
10.1007/s00234-001-0761-z
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The purpose of the study was the evaluation of 2-polyhydroxyethyl methacrylate (2-P-HEMA) for endovascular liquid embolization of experimental side-wall aneurysms following stent protection in a canine model. The swelling behaviour and polymerization characteristics of 2-P-HEMA in different solutions were investigated in vitro. Different methods for applications were tested in a latex aneurysm model under pulsatile flow conditions. Twenty broad-based carotid side-wall aneurysms were microsurgically produced in five dogs. Four weeks after surgery self-expandable nitinol stents were placed, covering the orifice of the aneurysms. 2-P-HEMA was injected via a microcatheter, which was positioned through the meshwork of the stent. Control angiography was performed immediately after treatment and after 1, 6 and 9 months. In-vivo stent placement succeeded in all but one case. Two aneurysms occluded spontaneously after stent placement. Combined embolization of 17 aneurysms using a stent and 2-P-HEMA was performed. Eleven aneurysms could be primarily completely occluded (65%). A small remaining neck was evident in six aneurysms. Efflux of 2-P-HEMA during the process of embolization was observed in seven aneurysms, due to an excess volume of 2-P-HEMA. The excessive 2-PHEMA led to significant vessel stenosis in two cases. Two carotid arteries (three treated aneurysms) occluded after 1 month, due to insufficient anticoagulation management. Histological examination of embolized aneurysms revealed no foreign-body or inflammatory reaction. A smooth neo-intimal layer covered the stented vessel segment. Liquid embolization of side-wall aneurysms with 2-P-HEMA is technically feasible. Embolotherapy of aneurysms with liquid agents still has the risk that embolic material will exit even when it is stent-protected. To avoid this problem, stents with smaller strut diameter and/or additional balloon-protection are required. The inert 2-P-HEMA seems to be a promising agent for combining techniques of aneurysm treatment.
引用
收藏
页码:503 / 512
页数:10
相关论文
共 44 条
[1]   BALLOON OCCLUSION OF THE INTERNAL CAROTID-ARTERY IN 40 CASES OF GIANT INTRACAVERNOUS ANEURYSM - TECHNICAL ASPECTS, CEREBRAL MONITORING, AND RESULTS [J].
ANON, VV ;
AYMARD, A ;
GOBIN, YP ;
CASASCO, A ;
RUFFENACHT, D ;
KHAYATA, MH ;
ABIZANDA, E ;
REDONDO, A ;
MERLAND, JJ .
NEURORADIOLOGY, 1992, 34 (03) :245-251
[2]  
BYRNE JV, 1994, NEUROL RES, V16, P425
[3]  
BYRNE JV, 1997, AM J NEURORADIOL, V41, P979
[4]   TISSUE INTERACTION WITH HYDROGEL SPONGES IMPLANTED IN THE RABBIT CORNEA [J].
CRAWFORD, GJ ;
CONSTABLE, IJ ;
CHIRILA, TV ;
VIJAYASEKARAN, S ;
THOMPSON, DE .
CORNEA, 1993, 12 (04) :348-357
[5]   OBLITERATION OF EXPERIMENTAL ANEURYSMS IN DOGS WITH ISOBUTYL-CYANOACRYLATE [J].
DEBRUN, GM ;
VARSOS, V ;
LISZCZAK, TM ;
DAVIS, KR ;
HEROS, RS ;
ZERVAS, NT .
JOURNAL OF NEUROSURGERY, 1984, 61 (01) :37-43
[7]  
GEREMIA G, 1995, AM J NEURORADIOL, V16, P1965
[8]   ANEURYSM RECURRENCE FOLLOWING ENDOVASCULAR BALLOON OCCLUSION [J].
HEILMAN, CB ;
KWAN, ESK ;
WU, JK .
JOURNAL OF NEUROSURGERY, 1992, 77 (02) :260-264
[9]   Intravascular stent and endovascular coil placement for a ruptured fusiform aneurysm of the basilar artery - Case report and review of the literature [J].
Higashida, RT ;
Smith, W ;
Gress, D ;
Urwin, R ;
Dowd, CF ;
Balousek, PA ;
Halbach, VV .
JOURNAL OF NEUROSURGERY, 1997, 87 (06) :944-949
[10]   Hydrogels in endovascular embolization .6. Toxicity tests of poly(2-hydroxyethyl methacrylate) particles on cell cultures [J].
Horak, D ;
Cervinka, M ;
Puza, V .
BIOMATERIALS, 1997, 18 (20) :1355-1359