Sensitive quantitative analysis of C-peptide in human plasma by 2-dimensional liquid chromatography-mass spectrometry isotope-dilution assay

被引:39
作者
Rogatsky, E
Balent, B
Goswami, G
Tomuta, V
Jayatillake, H
Cruikshank, G
Vele, L
Stein, DT
机构
[1] Albert Einstein Coll Med, Div Endocrinol & Metab, Dept Med, Bronx, NY 10461 USA
[2] Albert Einstein Coll Med, Gen Clin Res Ctr, Bronx, NY 10461 USA
[3] Yeshiva Univ, Bronx, NY 10033 USA
关键词
D O I
10.1373/clinchem.2005.063081
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Isotope-dilution assays (IDAs) are well established for quantification of metabolites or small drug molecules in biological fluids. Because of their increased specificity, IDAs are an alternative to immunoassays for measuring C-peptide. Methods: We evaluated a 2-dimensional liquid chromatography-mass spectrometry (2D LC/MS) IDA method. Sample preparation was by off-line solid-phase extraction, and C-peptide separation was performed on an Agilent 1100 2D LC system with a purification method based on high-pressure switching between 2 high-resolution reversed-phase columns. Because of the low fragmentation efficiency of C-peptide, multiple-reaction monitoring analysis was omitted and selective-ion monitoring mode was chosen for quantification. Native and isotope-labeled ([M+18] and [M+30]) C-peptides were monitored in the +3 state at m/z 1007.7, 1013.7, and 1017.7. Results: The assay was linear (r(2) = 0.9995), with a detection limit of 300 amole (1 pg) on column. Inter- and intraday CVs for C-peptide were <= 2%. Comparison with an established polyclonal-based RIA showed high correlation (r = 0.964). Plasma concentrations of total C-peptide measured by RIA were consistently higher than by IDA LC/MS, consistent with the higher specificity of IDAs compared with immunoassays. Conclusions: The 2D LC/MS IDA approach eliminates matrix effects, enhancing assay performance and reliability, and has a detection limit 100-fold lower than any previously reported LC/MS method. Isotope-labeled C-peptide(s) can be clearly differentiated from endogenous C-peptide by the difference in m/z ratio, so that both peptides can be quantified simultaneously. The method is highly precise, robust, and applicable to pharmacokinetic detection of plasma peptides. (c) 2006 American Association for Clinical Chemistry.
引用
收藏
页码:872 / 879
页数:8
相关论文
共 28 条
[1]   Accurate assessment of β-cell function -: The hyperbolic correction [J].
Bergman, RN ;
Ader, M ;
Huecking, K ;
Van Citters, G .
DIABETES, 2002, 51 :S212-S220
[2]   WHO INTERNATIONAL REFERENCE REAGENTS FOR HUMAN PROINSULIN AND HUMAN INSULIN C-PEPTIDE [J].
BRISTOW, AF ;
DAS, REG .
JOURNAL OF BIOLOGICAL STANDARDIZATION, 1988, 16 (03) :179-186
[3]   Assays for insulin, proinsulin(s) and C-peptide [J].
Clark, PM .
ANNALS OF CLINICAL BIOCHEMISTRY, 1999, 36 :541-564
[4]   A mass spectrometric method for quantitation of intact insulin in blood samples [J].
Darby, SM ;
Miller, ML ;
Allen, RO ;
LeBeau, M .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2001, 25 (01) :8-14
[5]   Quantitation of the large polypeptide glucagon by protein precipitation and LC/MS [J].
Delinsky, DC ;
Hill, KT ;
White, CA ;
Bartlett, MG .
BIOMEDICAL CHROMATOGRAPHY, 2004, 18 (09) :700-705
[6]   Quantitative determination of the polypeptide motilin in rat plasma by externally calibrated liquid chromatography/electrospray ionization mass spectrometry [J].
Delinsky, DC ;
Hill, KT ;
White, CA ;
Bartlett, MG .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (03) :293-298
[7]   HEPATIC REMOVAL OF INSULIN IN NORMAL MAN - DOSE-RESPONSE TO ENDOGENOUS INSULIN-SECRETION [J].
EATON, RP ;
ALLEN, RC ;
SCHADE, DS .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 1983, 56 (06) :1294-1300
[8]  
Fierens C, 2000, RAPID COMMUN MASS SP, V14, P936, DOI 10.1002/(SICI)1097-0231(20000530)14:10<936::AID-RCM960>3.0.CO
[9]  
2-A
[10]   Application of a C-peptide electrospray ionization-isotope dilution-liquid chromatography-tandem mass spectrometry measurement procedure for the evaluation of five C-peptide immunoassays for urine [J].
Fierens, C ;
Stöckl, D ;
Baetens, D ;
De Leenheer, AP ;
Thienpont, LM .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 792 (02) :249-259