Reconstruction of 6 MV photon spectra from measured transmission including maximum energy estimation

被引:25
作者
Baker, CR
Peck, KK
机构
[1] WALSGRAVE GEN HOSP,NHS TRUST,COVENTRY CV2 2DY,W MIDLANDS,ENGLAND
[2] UNIV BIRMINGHAM,DEPT PHYS & SPACE RES,BIRMINGHAM B15 2TT,W MIDLANDS,ENGLAND
关键词
D O I
10.1088/0031-9155/42/11/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Photon spectra from a nominally 6 MV beam under standard clinical conditions and at higher and lower beam qualities have been derived from narrow-beam transmission measurements using a previously published three-parameter reconstruction model. Estimates of the maximum photon energy present in each spectrum were derived using a reduced number of model parameters. An estimate of the maximum contribution of background, or room, scatter to transmission measurements has been made for this study and is shown to be negligible in terms of the quality index and percentage depth-dose of the derived spectra. Percentage depth-dose data for standard beam conditions derived from the reconstructed spectrum were found to agree with direct measurements to within approximately 1% for depths of up to 25 cm in water. Quality indices expressed in terms of TPR1020 for all spectra were found to agree with directly measured values to within 1%. The experimental procedure and reconstruction model are therefore shown to produce photon spectra whose derived quality indices and percentage depth-dose values agree with direct measurement to within expected experimental uncertainty.
引用
收藏
页码:2041 / 2051
页数:11
相关论文
共 29 条
[1]   STOPPING POWER DATA FOR HIGH-ENERGY PHOTON BEAMS [J].
ANDREO, P ;
BRAHME, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (08) :839-858
[2]  
ATTIX FH, 1986, INTRO RADIOBIOLOGICA, pCH10
[3]   RECONSTRUCTION OF MEGAVOLTAGE PHOTON SPECTRA BY ATTENUATION ANALYSIS [J].
BAKER, CR ;
AMAEE, B ;
SPYROU, NM .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (04) :529-542
[4]  
Baker CR, 1993, THESIS U SURREY
[5]  
BERGER MJ, 1907, XCOM PHOTON CROSS SE
[6]  
*BIR, 1983, BR J RADIOL S, V17
[7]   CONVERSION OF DEPTH DOSES FROM ONE FSD TO ANOTHER [J].
BURNS, JE .
BRITISH JOURNAL OF RADIOLOGY, 1958, 31 (371) :643-643
[8]   RECONSTRUCTION OF 12-MV BREMSSTRAHLUNG SPECTRA FROM MEASURED TRANSMISSION DATA BY DIRECT RESOLUTION OF THE NUMERIC SYSTEM AF=T [J].
CATALA, A ;
FRANCOIS, P ;
BONNET, J ;
SCOUARNEC, C .
MEDICAL PHYSICS, 1995, 22 (01) :3-10
[9]   FORWARD-DIRECTED BREMSSTRAHLUNG OF 10-MEV TO 30-MEV ELECTRONS INCIDENT ON THICK TARGETS OF AL AND PB [J].
FADDEGON, BA ;
ROSS, CK ;
ROGERS, DWO .
MEDICAL PHYSICS, 1990, 17 (05) :773-785
[10]  
FADDEGON BA, 1991, 0245 NRCC PIRS