Development and surgical implantation of a vision prosthesis model into the ovine eye

被引:18
作者
Kerdraon, YA
Downie, JA
Suaning, GJ
Capon, MR
Coroneo, MT
Lovell, NH [1 ]
机构
[1] Univ New S Wales, Grad Sch Biomed Engn, Sydney, NSW 2052, Australia
[2] Prince Wales Hosp, Dept Ophthalmol, Randwick, NSW, Australia
关键词
fixation sutures; intraocular; lensectomy; surgical implantation; vision prosthesis;
D O I
10.1046/j.1442-9071.2002.00485.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: A surgical technique was designed and tested to enable the implantation of an intraocular electrical retinal stimulator. Method: An inoperative perspex and silicone model was constructed to closely resemble the anticipated proper-ties of the proposed visual prosthesis. The animal model chosen for these experiments was the sheep, because the dimensions of its ocular anatomy are approximately 30% larger than the human's, being otherwise grossly similar. The surgical method involved transplanar port-hole lensectomy and vitrectomy, insertion of the model implant through a limbal incision, and fixation of the perspex subunit close to the location of the native crystalline lens, by way of trans-pars plana fixation sutures. Adequate preretinal positioning of the implant's silicone extension was obtained by way of its inherent elastic recoil. Results: The procedure was performed without macroscopic evidence of undue surgical trauma, Conclusion: Although further long-term experiments are required to fully assess the surgical procedure and biocompatibility of the implant, intraoperative assessment and postmortem computed tomographic imaging of the globe has confirmed the successful intraocular positioning and fixation of the implant.
引用
收藏
页码:36 / 40
页数:5
相关论文
共 8 条
[1]  
Berler D K, 1991, Trans Am Ophthalmol Soc, V89, P215
[2]  
DONOVAN RH, 1974, ANN OPHTHALMOL, V6, P1175
[3]  
Humayun MS, 1999, INVEST OPHTH VIS SCI, V40, P143
[4]   ANTERIOR SEGMENT VASCULAR CASTING [J].
OLVER, JM ;
MCCARTNEY, ACE .
EYE, 1989, 3 :302-307
[5]  
Samuelson DA, 1999, VET OPHTHALMOLOGY
[6]   MORPHOMETRIC ANALYSIS OF MACULAR PHOTORECEPTORS AND GANGLION-CELLS IN RETINAS WITH RETINITIS-PIGMENTOSA [J].
STONE, JL ;
BARLOW, WE ;
HUMAYUN, MS ;
DEJUAN, E ;
MILAM, AH .
ARCHIVES OF OPHTHALMOLOGY, 1992, 110 (11) :1634-1639
[7]   CMOS neurostimulation ASIC with 100 channels, scaleable output, and bidirectional radio-frequency telemetry [J].
Suaning, GJ ;
Lovell, NH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (02) :248-260
[8]   The bionic eye (electronic visual prosthesis): A review [J].
Suaning, GJ ;
Lovell, NH ;
Schindhelm, K ;
Coroneo, MT .
AUSTRALIAN AND NEW ZEALAND JOURNAL OF OPHTHALMOLOGY, 1998, 26 (03) :195-202