Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit

被引:46
作者
Seguin, P
Le Rouzo, A
Tanguy, M
Guillou, YM
Feuillu, A
Mallédant, Y
机构
[1] CHRU Pontchaillou, Serv Reanimat Chirurg, F-35035 Rennes, France
[2] CHRU Pontchaillou, Biochim Lab, F-35035 Rennes, France
关键词
pulse oximetry; mechanical ventilation; monitoring; critical care;
D O I
10.1097/00003246-200003000-00017
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To compare pulse oximetry saturation (SpO(2)) with arterial blood gas saturation (SaO(2)) obtained during clinical routine to determine the optimal lowest reliable value of SpO(2) in ventilator-dependent patients before setting up a nurse-directed protocol of FIO2 titration. Design: Prospective clinical study. Setting: Surgical intensive care unit in a university hospital. Patients: Thirty-three patients with a pulse oximeter probe in whom arterial blood gas was measured with a radial artery line. Interventions: SpO(2) was recorded by the nurses and compared with SaO(2) obtained by blood gas analysis with a co-oximeter. Two sensors currently used in our surgical intensive care unit and connected to a monitor (HP OmniCare M1165/66A; Hewett Packard, Andover, MA) were tested. In group I, the Durensor DS 100A (Nellcor Puritan Bennett, Pleasanton, CA), a reusable sensor, was used. In group II, the Oxisensor D25L (Nellcor Puritan Bennett), a nonreusable sensor, was used. Measurements and Main Results: In group I, 64 data pairs were obtained. In this group, SaO(2) ranged from 87 to 98% and SpO(2) ranged from 92 to 100%, The bias was -1.90% and the limits of agreement ranged from -5.56 to 1.76%. In group 11, 47 data pairs were obtained. In this group, SaO(2) ranged from 87 to 99% and SpO(2) ranged from 92 to 100%. The bias was -2.49% and the limits of agreement ranged from -6.62 to 1.64%. Conclusions: In the range of SaO(2) tested, regardless of the sensor used, SpO(2) overestimated SaO(2). Large limits of agreement were found. Based on this result, the authors concluded that before defining a nurse-directed protocol of FIO2 titration with SpO(2), the material used daily must be evaluated. A minimum threshold SpO(2) value of 96% in both groups I and II is more reliable to ensure SaO(2) greater than or equal to 90%.
引用
收藏
页码:703 / 706
页数:4
相关论文
共 16 条
[1]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[2]   THE EFFECT OF CARBON-MONOXIDE INHALATION ON PULSE OXIMETRY AND TRANS-CUTANEOUS PO2 [J].
BARKER, SJ ;
TREMPER, KK .
ANESTHESIOLOGY, 1987, 66 (05) :677-679
[3]   PULSE OXIMETRY IN THE POSTOPERATIVE CARE OF CARDIAC SURGICAL PATIENTS - A RANDOMIZED CONTROLLED TRIAL [J].
BIERMAN, MI ;
STEIN, KL ;
SNYDER, JV .
CHEST, 1992, 102 (05) :1367-1370
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   A COMPARISON OF THE PERFORMANCE OF 20 PULSE OXIMETERS UNDER CONDITIONS OF POOR PERFUSION [J].
CLAYTON, DG ;
WEBB, RK ;
RALSTON, AC ;
DUTHIE, D ;
RUNCIMAN, WB .
ANAESTHESIA, 1991, 46 (01) :3-10
[6]   THE ACCURACY OF THE BIOX 3700 PULSE OXIMETER IN PATIENTS RECEIVING VASOACTIVE THERAPY [J].
IBANEZ, J ;
VELASCO, J ;
RAURICH, JM .
INTENSIVE CARE MEDICINE, 1991, 17 (08) :484-486
[7]   RELIABILITY OF PULSE OXIMETRY IN TITRATING SUPPLEMENTAL OXYGEN-THERAPY IN VENTILATOR-DEPENDENT PATIENTS [J].
JUBRAN, A ;
TOBIN, MJ .
CHEST, 1990, 97 (06) :1420-1425
[8]   RANDOMIZED EVALUATION OF PULSE OXIMETRY IN 20,802 PATIENTS .2. PERIOPERATIVE EVENTS AND POSTOPERATIVE COMPLICATIONS [J].
MOLLER, JT ;
JOHANNESSEN, NW ;
ESPERSEN, K ;
RAVLO, O ;
PEDERSEN, BD ;
JENSEN, PF ;
RASMUSSEN, NH ;
RASMUSSEN, LS ;
PEDERSEN, T ;
COOPER, JB ;
GRAVENSTEIN, JS ;
CHRAEMMERJORGENSEN, B ;
DJERNES, M ;
WIBERGJORGENSEN, F ;
HESLET, L ;
JOHANSEN, SH .
ANESTHESIOLOGY, 1993, 78 (03) :445-453
[9]   RANDOMIZED EVALUATION OF PULSE OXIMETRY IN 20,802 PATIENTS .1. DESIGN, DEMOGRAPHY, PULSE OXIMETRY FAILURE RATE, AND OVERALL COMPLICATION RATE [J].
MOLLER, JT ;
PEDERSEN, T ;
RASMUSSEN, LS ;
JENSEN, PF ;
PEDERSEN, BD ;
RAVLO, O ;
RASMUSSEN, NH ;
ESPERSEN, K ;
JOHANNESSEN, NW ;
COOPER, JB ;
GRAVENSTEIN, JS ;
CHRAEMMERJORGENSEN, B ;
WIBERGJORGENSEN, F ;
DJERNES, M ;
HESLET, L ;
JOHANSEN, SH .
ANESTHESIOLOGY, 1993, 78 (03) :436-444
[10]   Uses and abuses of pulse oximetry [J].
Moyle, JTB .
ARCHIVES OF DISEASE IN CHILDHOOD, 1996, 74 (01) :77-80