A compact small-beam XRF instrument for in-situ analysis of objects of historical and/or artistic value

被引:60
作者
Vittiglio, G
Janssens, K
Vekemans, B
Adams, F
Oost, A
机构
[1] Univ Instelling Antwerp, Dept Chem, Ctr Micro & Trace Anal, B-2610 Antwerp, Belgium
[2] Art Hist Musea City Antwerp, Museum Vleeshuis, Antwerp, Belgium
关键词
X-ray fluorescence analysis; microanalysis; non-destructive analysis; archaeological bronze; quantitative analysis; corrosion of bronze; micro-polishing;
D O I
10.1016/S0584-8547(99)00100-7
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The analytical characteristics, possibilities and limitations of a compact and easily transportable small-beam XRF instrument are described. The instrument consists of a compact, mini-focus Mo X-ray tube that is collimated to produce a sub-mm beam and a peltier-cooled PIN diode detector. Relative MDLs in highly scattering matrices are situated in the 10-100-ppm range; for metallic matrices featuring strong matrix lines, the MDLs of the instrument are approximately a factor 2 higher. Since only a small irradiation area is required, a simple micro-polishing technique that may be performed in situ in combination with the measurements is shown to be effective for the determination of the bulk composition of corroded bronze objects. As an example, a series of Egyptian bronze objects date from XXIInd Egyptian Dynasty (ca. 1090 Be) to the Roman era (30 BC to 640 AD) was analyzed in order to contribute to the very limited database on Cu-alloy compositions from this period. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1697 / 1710
页数:14
相关论文
共 19 条
[1]   Micro and surface analysis in art and archaeology - Plenary lecture [J].
Adams, F ;
Adriaens, A ;
Aerts, A ;
DeRaedt, I ;
Janssens, K ;
Schalm, O .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (03) :257-265
[2]  
Cowell M.R., 1986, SCI EGYPTOLOGY, P463
[3]   IRON IN ANCIENT COPPER [J].
CRADDOCK, PT ;
MEEKS, ND .
ARCHAEOMETRY, 1987, 29 :187-204
[4]   GENERAL-APPROACH FOR QUANTITATIVE ENERGY DISPERSIVE-X-RAY FLUORESCENCE ANALYSIS BASED ON FUNDAMENTAL PARAMETERS [J].
HE, F ;
VANESPEN, PJ .
ANALYTICAL CHEMISTRY, 1991, 63 (20) :2237-2244
[5]   A micro-XRF spectrometer based on a rotating anode generator and capillary optics [J].
Janssens, K ;
Vekemans, B ;
Vincze, L ;
Adams, F ;
Rindby, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (13) :1661-1678
[6]  
JOHANSSON SA, 1988, PIXE NOVEL TECHNIQUE, P244
[7]  
LACOUDRE N, 1997, RX9716022 EDF VAL
[8]   STUDY AND CONSERVATION OF MUSEUM OBJECTS - USE OF CLASSICAL ANALYTICAL TECHNIQUES [J].
LAHANIER, C ;
PREUSSER, FD ;
VANZELST, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1986, 14 (01) :1-9
[9]  
Lucas A., 1948, Ancient Egyptian Materials Industries, Vfourth
[10]  
MALMQVIST KG, 1995, PARTICLE INDUCES XRA, pCH7