Precise temperature control in microfluidic devices using Joule heating of ionic liquids

被引:85
作者
de Mello, AJ [1 ]
Habgood, M [1 ]
Lancaster, NL [1 ]
Welton, T [1 ]
Wootton, RCR [1 ]
机构
[1] Univ London Imperial Coll Sci Technol & Med, London SW7 2AZ, England
关键词
D O I
10.1039/b405760k
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Microfluidic devices for spatially localised heating of microchannel environments were designed, fabricated and tested. The devices are simple to implement, do not require complex manufacturing steps and enable intrachannel temperature control to within +/-0.2 degreesC. Ionic liquids held in co-running channels are Joule heated with an a. c. current. The nature of the devices means that the internal temperature can be directly assessed in a facile manner.
引用
收藏
页码:417 / 419
页数:3
相关论文
共 20 条
[1]   Flammability, thermal stability, and phase change characteristics of several trialkylimidazolium salts [J].
Fox, DM ;
Awad, WH ;
Gilman, JW ;
Maupin, PH ;
De Long, HC ;
Trulove, PC .
GREEN CHEMISTRY, 2003, 5 (06) :724-727
[2]   New developments in catalysis using ionic liquids [J].
Gordon, CM .
APPLIED CATALYSIS A-GENERAL, 2001, 222 (1-2) :101-117
[3]   Chemical and physical processes for integrated temperature control in microfluidic devices [J].
Guijt, RM ;
Dodge, A ;
van Dedem, GWK ;
de Rooij, NF ;
Verpoorte, E .
LAB ON A CHIP, 2003, 3 (01) :1-4
[4]  
Holbrey JD., 1999, Clean. Prod. Process., V1, P223, DOI [10.1007/s100980050036, DOI 10.1007/S100980050036]
[5]   Characterization and comparison of hydrophilic and hydrophobic room temperature ionic liquids incorporating the imidazolium cation [J].
Huddleston, JG ;
Visser, AE ;
Reichert, WM ;
Willauer, HD ;
Broker, GA ;
Rogers, RD .
GREEN CHEMISTRY, 2001, 3 (04) :156-164
[6]   Non-conventional electrolytes for electrochemical applications [J].
Ito, Y ;
Nohira, T .
ELECTROCHIMICA ACTA, 2000, 45 (15-16) :2611-2622
[7]   Chemistry in microstructured reactors [J].
Jähnisch, K ;
Hessel, V ;
Löwe, H ;
Baerns, M .
ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2004, 43 (04) :406-446
[8]   Miniaturized total analysis systems for biological analysis [J].
Jakeway, SC ;
de Mello, AJ ;
Russell, EL .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 366 (6-7) :525-539
[9]   Integrated system for rapid PCR-based DNA analysis in microfluidic devices [J].
Khandurina, J ;
McKnight, TE ;
Jacobson, SC ;
Waters, LC ;
Foote, RS ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 2000, 72 (13) :2995-3000
[10]   Chemical amplification: Continuous-flow PCR on a chip [J].
Kopp, MU ;
de Mello, AJ ;
Manz, A .
SCIENCE, 1998, 280 (5366) :1046-1048