The influence of head and neck geometry on stability of total hip replacement - A mechanical test study

被引:53
作者
Bader, R [1 ]
Scholz, R
Steinhauser, E
Zimmermann, S
Busch, R
Mittelmeier, W
机构
[1] Tech Univ Munich, Klin Orthopad & Sportorthopad, D-8000 Munich, Germany
[2] Univ Leipzig, Orthopad Klin & Poliklin, Leipzig, Germany
[3] Tech Univ Munich, Inst Med Stat & Epidemiol, D-8000 Munich, Germany
[4] Univ Rostock, Orthopad Klin & Poliklin, D-2500 Rostock, Germany
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 2004年 / 75卷 / 04期
关键词
D O I
10.1080/00016470410001178-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Dislocation after replacement may be caused by poor implant design or positioning, or by the surgical approach taken. We evaluated the influence of head and neck design on range of motion and stability (with respect to risk of dislocation) in total hip endoprostheses. Material and methods Using a test device, we determined the stability afforded by different head sizes and neck geometries for various implant positions. Results Increasing head diameter led to an enhancement of range of motion as well as resistance against subluxation, and thus to improved stability in any movement combination and implant orientation. Smaller femoral heads were associated with increased risk of dislocation, especially in a poor implant position such as retroversion, and steep positioning of the cup. Skirted metal or mushroom-shaped ceramic heads had a reduced range of motion until impingement of approx. 20degrees, as compared to spherical standard heads. Furthermore, after identical joint loading, skirted heads dislocated more readily than standard heads with corresponding diameters. Interpretation To obtain sufficient joint mobility and stability, neck geometry and implant position should be considered when choosing the femoral head size.
引用
收藏
页码:415 / 421
页数:7
相关论文
共 24 条
[1]
Method for the evaluation of factors influencing the dislocation stability of total hip endoprotheses [J].
Bader, R ;
Scholz, R ;
Steinhauser, E ;
Busch, R ;
Mittlemeier, W .
BIOMEDIZINISCHE TECHNIK, 2004, 49 (05) :137-144
[2]
Computer-based motion simulation of total hip prostheses with ceramic-on-ceramic wear couple. Analysis of implant design and orientation as influence parameters [J].
Bader, R ;
Steinhauser, E ;
Gradinger, R ;
Willmann, G ;
Mittelmeier, W .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2002, 140 (03) :310-316
[3]
The effect of femoral component head size on posterior dislocation of the artificial hip joint [J].
Bartz, RL ;
Noble, PC ;
Kadakia, NR ;
Tullos, HS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (09) :1300-1307
[4]
Jumbo femoral head for the treatment of recurrent dislocation following total hip replacement [J].
Beaule, PE ;
Schmalzried, TP ;
Udomkiat, P ;
Amstutz, HC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (02) :256-263
[5]
Femoral head sizes larger than 32 mm against highly cross-linked polyethylene [J].
Burroughs, BR ;
Rubash, HE ;
Harris, WH .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (405) :150-157
[6]
Femoral head size is a risk factor for total hip luxation -: A study of 42,987 primary hip arthroplasties from the Norwegian Arthroplasty Register [J].
Byström, S ;
Espehaug, B ;
Furnes, O ;
Havelin, LI .
ACTA ORTHOPAEDICA SCANDINAVICA, 2003, 74 (05) :514-524
[7]
Callaghan JJ, 2001, CLIN ORTHOP RELAT R, P157
[8]
CHARNLEY J, 1973, CLIN ORTHOP RELAT R, P9
[9]
GACHTER A, 1989, ORTHOPADE, V18, P533
[10]
Genoud P, 2000, P 12 ESB C DUBL, P137