Experimental investigation of a fast Monte Carlo photon beam dose calculation algorithm

被引:44
作者
Fippel, M [1 ]
Laub, W [1 ]
Huber, B [1 ]
Nüsslin, F [1 ]
机构
[1] Univ Tubingen, Radiol Klin, Abt Med Phys, D-72076 Tubingen, Germany
关键词
D O I
10.1088/0031-9155/44/12/313
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An experimental verification of the recently developed XVMC code, a fast Monte Carlo algorithm to calculate dose distributions of photon beams in treatment planning, is presented. The treatment head is modelled by a point source with energy distribution (primary photons) and an additional head scatter contribution. Utility software is presented, allowing the determination of the parameters for this model using a single measured depth dose curve in water. The simple beam model is considered to be a starting point for more complex models being planned for future versions of the code. This paper is mainly focused on the influence of the different techniques on variance reduction and material properly determination for dose distributions. It is demonstrated that XVMC and the simple beam model reproduce measured (by a diamond detector) relative dose distributions with an accuracy of better than 12% in various homogeneous and inhomogeneous phantoms. Furthermore, relative dose distributions in solid state phantoms have been measured by film. Also for these cases, measured and calculated dose distributions agree within experimental uncertainty. The short calculation time (depending on voxel resolution, statistical accuracy, field size and energy, a span of 1 min to 1 h using a present-day personal computer) and an interface to a commercial planning system will allow the implementation of the code for routine treatment planning of clinical electron and photon beams.
引用
收藏
页码:3039 / 3054
页数:16
相关论文
共 30 条
[21]   EXPERIMENTAL EVALUATION OF A 2D AND 3D ELECTRON PENCIL BEAM ALGORITHM [J].
MAH, E ;
ANTOLAK, J ;
SCRIMGER, JW ;
BATTISTA, JJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (09) :1179-1194
[22]   DIFFERENTIAL PENCIL BEAM DOSE COMPUTATION MODEL FOR PHOTONS [J].
MOHAN, R ;
CHUI, C ;
LIDOFSKY, L .
MEDICAL PHYSICS, 1986, 13 (01) :64-73
[23]  
Mohan R, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P16
[24]  
NAHUM AE, 1997, P CHALL CONF RAD NIC, P1
[25]  
Nelson W., 1985, SLAC265
[26]   A MACRO MONTE-CARLO METHOD FOR ELECTRON-BEAM DOSE CALCULATIONS [J].
NEUENSCHWANDER, H ;
BORN, EJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :107-125
[27]   BEAM - A MONTE-CARLO CODE TO SIMULATE RADIOTHERAPY TREATMENT UNITS [J].
ROGERS, DWO ;
FADDEGON, BA ;
DING, GX ;
MA, CM ;
WE, J ;
MACKIE, TR .
MEDICAL PHYSICS, 1995, 22 (05) :503-524
[28]   ELECTRON-BEAM DOSE DISTRIBUTIONS NEAR STANDARD INHOMOGENEITIES [J].
SHORTT, KR ;
ROSS, CK ;
BIELAJEW, AF ;
ROGERS, DWO .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (03) :235-249
[29]  
SIANTAR CLH, 1997, UCRLJC126732 LLNL
[30]   A patient-specific Monte Carlo dose-calculation method for photon beams [J].
Wang, L ;
Chui, CS ;
Lovelock, M .
MEDICAL PHYSICS, 1998, 25 (06) :867-878