Electrochemical studies of zinc in zinc-insulin solution

被引:11
作者
Barbosa, RM
Rosario, LM
Brett, CMA
Brett, AMO
机构
[1] UNIV COIMBRA,DEPT QUIM,FAC CIENCIAS & TECNOL,P-3049 COIMBRA,PORTUGAL
[2] UNIV COIMBRA,DEPT BIOQUIM,FAC CIENCIAS & TECNOL,P-3049 COIMBRA,PORTUGAL
[3] UNIV COIMBRA,CTR NEUROCIENCIAS COIMBRA,P-3049 COIMBRA,PORTUGAL
[4] UNIV COIMBRA,FAC FARM,LAB METEDOS INSTRUMENTAIS ANAL,P-3049 COIMBRA,PORTUGAL
关键词
zinc; insulin; square-wave anodic stripping voltammetry; microelectrodes; batch injection analysis;
D O I
10.1039/an9962101789
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The electrochemical determination of zinc arising from zinc-insulin complexes was investigated and it was demonstrated that zinc in zinc-insulin solution can be measured in the presence of dissolved oxygen by square-wave anodic stripping voltammetry (SWASV) at mercury thin-film electrodes on glassy carbon disc minielectrode and cylindrical carbon fibre microelectrode substrates. Reoxidation signals arise from complexed zinc at low insulin concentrations (<100 nmol l(-1)) and from labile zinc at higher concentrations; the latter can be quantified through linear calibration curves, Batch injection analysis with SWASV was successfully tested for the determination of zinc in zinc-insulin solutions in small sample volumes. Since intracellularly stored insulin exists in the form of a zinc-insulin complex, these techniques are very promising for the indirect study of insulin release from pancreatic beta-cells.
引用
收藏
页码:1789 / 1793
页数:5
相关论文
共 26 条
  • [1] BLUNDELL T, 1972, ADV PROTEIN CHEM, P325
  • [2] Blundell T, 1972, ADV PROTEIN CHEM, V1972, P279, DOI 10.1016/S0065-3233(08)60143-6
  • [3] Anodic stripping voltammetry of trace metals by batch injection analysis
    Brett, CMA
    Brett, AMO
    Tugulea, L
    [J]. ANALYTICA CHIMICA ACTA, 1996, 322 (03) : 151 - 157
  • [4] AMPEROMETRIC AND VOLTAMMETRIC DETECTION IN BATCH INJECTION-ANALYSIS
    BRETT, CMA
    BRETT, AMO
    MITOSERIU, LC
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (19) : 3145 - 3150
  • [5] AMPEROMETRIC BATCH INJECTION-ANALYSIS - THEORETICAL ASPECTS OF CURRENT TRANSIENTS AND COMPARISON WITH WALL-JET ELECTRODES IN CONTINUOUS-FLOW
    BRETT, CMA
    BRETT, AMO
    MITOSERIU, LC
    [J]. ELECTROANALYSIS, 1995, 7 (03) : 225 - 229
  • [6] FLOW-INJECTION AMPEROMETRIC DETERMINATION OF INSULIN BASED UPON ITS OXIDATION AT A MODIFIED ELECTRODE
    COX, JA
    GRAY, TJ
    [J]. ANALYTICAL CHEMISTRY, 1989, 61 (21) : 2462 - 2464
  • [7] DEREWENDA U, 1990, INSULIN, pCH2
  • [8] EFENDIC S, 1994, DIABETES METAB, V20, P81
  • [9] EFFECTS OF ZN-2+ ON GLUCOSE-INDUCED ELECTRICAL-ACTIVITY AND INSULIN RELEASE FROM MOUSE PANCREATIC-ISLETS
    FERRER, R
    SORIA, B
    DAWSON, CM
    ATWATER, I
    ROJAS, E
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1984, 246 (05): : C520 - C527
  • [10] IMMUNOASSAY OF INSULIN WITH INSULIN-ANTIBODY PRECIPITATE
    HALES, CN
    RANDLE, PJ
    [J]. BIOCHEMICAL JOURNAL, 1963, 88 (01) : 137 - &