AMPEROMETRIC BATCH INJECTION-ANALYSIS - THEORETICAL ASPECTS OF CURRENT TRANSIENTS AND COMPARISON WITH WALL-JET ELECTRODES IN CONTINUOUS-FLOW

被引:60
作者
BRETT, CMA [1 ]
BRETT, AMO [1 ]
MITOSERIU, LC [1 ]
机构
[1] UNIV BUCHAREST,DEPT BIOPHYS,BUCHAREST,ROMANIA
关键词
BATCH INJECTION ANALYSIS; WALL JET ELECTRODE; AUTOMATED ANALYSIS; HYDRODYNAMIC AMPEROMETRY;
D O I
10.1002/elan.1140070305
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Amperometric detection using the batch injection analysis technique has been studied and compared with the current obtained at the wall-jet electrode under similar conditions using the oxidation of potassium ferrocyanide and the injection of samples of volume 10 to 100 mu L. The form of the current transients has been analyzed and compared with the steady-state response at wail-jet disk electrodes. Good agreement is found at high dispension rate, differences at lower dispension rates being ascribed to radial diffusion effects. The importance of tip-electrode distance and injection volume have been evaluated; it is shown that the optimum distance is 3 mm and that maximum sensitivity requires a minimum injection volume of 14 mu L. A detection limit of 50 mu M using simple amperometric detection is estimated. Implications of these results for the efficient application of amperometric batch injection analysis are discussed.
引用
收藏
页码:225 / 229
页数:5
相关论文