Flat-panel X-ray detector using amorphous silicon technology - Reduced radiation dose for the detection of foreign bodies

被引:39
作者
Volk, M [1 ]
Strotzer, M [1 ]
Gmeinwieser, J [1 ]
Alexander, J [1 ]
Frund, R [1 ]
Seitz, J [1 ]
Manke, C [1 ]
Spahn, M [1 ]
Feuerbach, S [1 ]
机构
[1] SIEMENS AG,MED ENGN,D-8520 ERLANGEN,GERMANY
关键词
amorphous silicon; digital imaging; foreign bodies;
D O I
10.1097/00004424-199707000-00001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. The authors evaluate a new flat-panel x-ray detector (FD) with respect to foreign body detection and reduction of radiation dose compared with screen-film radiography. METHODS. Flat-panel x-ray detector is based on amorphous silicon technology and uses a 1 k x 1 k photo-detector matrix with a poxel size of 143 x 143 mu m and 12-bit digital output. A thallium-doted cesiu iodide scintillation layer converts x-rays into light. A ex vivo experimental model was used to determine the detectability of foreign bodies. Foreign bodies with varying sizes were examined: glass with and without addition of lead, bone, aluminium, iron, copper, gravel fragments, and graphite. Four hundred observation fields were examined using conventional radiography (speed, 400; system dose: 2.5 mu Gy) as well as FD with a simulated speed of 400, 800, 1200, and 1600, corresponding to a detector dose of 2.5 mu Gy, 1.25 mu Gy, 0.87 mu Gy, and 0.625 mu Gy, respectively. Four independent radiologists performed receiver operating characteristic analysis of 8000 observation. RESULTS. Flat-panel x-ray detector with a simulatio nspeed of 400 was significantly superior (P = 0.012) to screen-film radiography (speed, 400). At a simulated speed of 800 and 1200 FD yielded results equivalent to screen-film radiography. Flat-panel x-ray detector was significantly inferior to screen-film radiography at a simulated speed of 1600 (P = 0.012). CONCLUSIONS. Flat-panel x-ray detector technology allows significant reduction in radiation dose compared with screen-film radiography without loss of diagnostic accuracy.
引用
收藏
页码:373 / 377
页数:5
相关论文
共 16 条
[1]   DIAGNOSIS AND TREATMENT OF RETAINED FOREIGN-BODIES IN THE HAND [J].
ANDERSON, MA ;
NEWMEYER, WL ;
KILGORE, ES .
AMERICAN JOURNAL OF SURGERY, 1982, 144 (01) :63-67
[2]   DEMONSTRATION OF MEGAVOLTAGE AND DIAGNOSTIC-X-RAY IMAGING WITH HYDROGENATED AMORPHOUS-SILICON ARRAYS [J].
ANTONUK, LE ;
BOUDRY, J ;
HUANG, WD ;
MCSHAN, DL ;
MORTON, EJ ;
YORKSTON, J ;
LONGO, MJ ;
STREET, RA .
MEDICAL PHYSICS, 1992, 19 (06) :1455-1466
[3]   A REAL-TIME, FLAT-PANEL, AMORPHOUS-SILICON, DIGITAL X-RAY IMAGER [J].
ANTONUK, LE ;
YORKSTON, J ;
HUANG, WD ;
SIEWERDSEN, JH ;
BOUDRY, JM ;
ELMOHRI, Y ;
MARX, MV .
RADIOGRAPHICS, 1995, 15 (04) :993-1000
[4]   DIGITAL SKELETAL RADIOGRAPHY [J].
BUCKWALTER, KA ;
BRAUNSTEIN, EM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 158 (05) :1071-1080
[5]   PENETRATING INJURIES - HOW EASY IS IT TO SEE GLASS (AND PLASTIC) ON RADIOGRAPHS [J].
DELACEY, G ;
EVANS, R ;
SANDIN, B .
BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (685) :27-30
[6]   HIGH-RESOLUTION COMPUTER DISPLAY OF PORTABLE, DIGITAL, CHEST RADIOGRAPHS OF ADULTS - SUITABILITY FOR PRIMARY INTERPRETATION [J].
FRANK, MS ;
JOST, RG ;
MOLINA, PL ;
ANDERSON, DJ ;
SOLOMON, SL ;
WHITMAN, RA ;
MOORE, SM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (03) :473-477
[7]  
HAUS AG, 1996, RADIOGRAPHICS, V5, P1165
[8]  
METZ CE, 1986, INVEST RADIOL, V21, P720, DOI 10.1097/00004424-198609000-00009
[10]   COMPUTED RADIOGRAPHY IN MUSCULOSKELETAL IMAGING - STATE-OF-THE-ART [J].
MURPHEY, MD ;
QUALE, JL ;
MARTIN, NL ;
BRAMBLE, JM ;
COOK, LT ;
DWYER, SJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 158 (01) :19-27