Improvement of evaporative light scattering detection of ceramides using triethylamine and formic acid in non-aqueous reversed phase liquid chromatography

被引:11
作者
Gaudin, K [1 ]
Chaminade, P [1 ]
Ferrier, D [1 ]
Baillet, A [1 ]
机构
[1] Univ Paris 11, Fac Pharm, Chim Analyt Lab, F-92290 Chatenay Malabry, France
关键词
D O I
10.1081/JLC-100101458
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The use of Porous Graphitized Carbon (PGC) stationary phase has enabled to emphasize the efficiency of the couple triethylamine (TEA) and formic acid (HCOOH) as a response enhancer for the Evaporative Light Scattering Detection (ELSD) of ceramides in Non-Aqueous Reversed Phase (NARP) Liquid Chromatography. 0.1% v/v of TEA and a stoechiometric amount of HCOOH greatly increase the intensity of the detection response whatever the nature of the eluent and the mobile phase composition was. This increase was not proportional to the amount of the TEA and HCOOH added as the response remained stable above 0.1 % v/v (7 mM).
引用
收藏
页码:387 / 397
页数:11
相关论文
共 22 条
[21]  
UNGER KK, 1983, ANAL CHEM, V55, P361
[22]  
Yano M, 1998, J LIPID RES, V39, P2091