Stereotactic transcranial magnetic stimulation: Correlation with direct electrical cortical stimulation

被引:94
作者
Krings, T
Buchbinder, BR
Butler, WE
Chiappa, KH
Jiang, HJ
Rosen, BR
Cosgrove, GR
机构
[1] HARVARD UNIV, MASSACHUSETTS GEN HOSP,SCH MED,DEPT NEUROSURG, DIV NEURORADIOL,CLIN NEUROPHYSIOL LAB, BOSTON, MA 02114 USA
[2] HARVARD UNIV, MASSACHUSETTS GEN HOSP,SCH MED,DEPT RADIOL, MGH NMR CTR, BOSTON, MA 02114 USA
关键词
cortical mapping; frameless stereotaxy; presurgical planning; transcranial magnetic stimulation;
D O I
10.1097/00006123-199712000-00016
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To evaluate stereotactic transcranial magnetic stimulation (TMS) as a tool for presurgical functional mapping of human motor cortex. METHODS: Transcranial magnetic stimulation using a frameless stereotactic system was performed in two patients with tumors near the central sulcus. TMS motor function maps were plotted on the patients' three-dimensional volumetric magnetic resonance imaging data and compared with direct electrical cortical stimulation at surgery with the patient under local anesthesia. RESULTS: Stereotactic TMS was well tolerated by both patients and was consistent with known somatotopic representation of human motor cortex. The results demonstrated a good correlation between the TMS and electrical cortical stimulation maps, with all TMS responses eliciting more than 75% of the maximum motor evoked potential falling within 1 cm of the electrical cortical stimulation site. CONCLUSIONS: Our findings indicate that stereotactic TMS is feasible and can provide accurate noninvasive localization of cortical motor function. It may prove to be a useful method for presurgical planning.
引用
收藏
页码:1319 / 1325
页数:7
相关论文
共 20 条
[1]  
BARKER AT, 1985, LANCET, V1, P1106
[2]   TOPOGRAPHIC MAPPING OF THE HUMAN MOTOR CORTEX WITH MAGNETIC STIMULATION - FACTORS AFFECTING ACCURACY AND REPRODUCIBILITY [J].
BRASILNETO, JP ;
MCSHANE, LM ;
FUHR, P ;
HALLETT, M ;
COHEN, LG .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1992, 85 (01) :9-16
[3]   OPTIMAL FOCAL TRANSCRANIAL MAGNETIC ACTIVATION OF THE HUMAN MOTOR CORTEX - EFFECTS OF COIL ORIENTATION, SHAPE OF THE INDUCED CURRENT PULSE, AND STIMULUS-INTENSITY [J].
BRASILNETO, JP ;
COHEN, LG ;
PANIZZA, M ;
NILSSON, J ;
ROTH, BJ ;
HALLETT, M .
JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 1992, 9 (01) :132-136
[4]  
BRIDGERS SL, 1991, ELECTROEN CLIN NEURO, P170
[5]   DEVELOPING A MORE FOCAL MAGNETIC STIMULATOR .1. SOME BASIC PRINCIPLES [J].
COHEN, D ;
CUFFIN, BN .
JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 1991, 8 (01) :102-111
[6]   EFFECTS OF COIL DESIGN ON DELIVERY OF FOCAL MAGNETIC STIMULATION - TECHNICAL CONSIDERATIONS [J].
COHEN, LG ;
ROTH, BJ ;
NILSSON, J ;
DANG, N ;
PANIZZA, M ;
BANDINELLI, S ;
FRIAUF, W ;
HALLETT, M .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1990, 75 (04) :350-357
[7]  
COHEN LG, 1991, ELECTROEN CLIN NEURO, P36
[8]   COMPUTER MODELING OF BRAIN CORTEX EXCITATION BY MAGNETIC-FIELD PULSES [J].
DELEO, R ;
CERRI, G ;
BALDUCCI, D ;
MOGLIE, F ;
SCARPINO, O ;
GUIDI, M .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1992, 16 (04) :149-156
[9]   ELECTRIC-FIELD INDUCED IN A SPHERICAL VOLUME CONDUCTOR FROM ARBITRARY COILS - APPLICATION TO MAGNETIC STIMULATION AND MEG [J].
EATON, H .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1992, 30 (04) :433-440
[10]  
HESS CW, 1986, J PHYSIOL-LONDON, V378, pP37