PURPOSE: To assess the retrievability of the G2 inferior vena cava (IVC) filter and factors influencing the safety and technical success of retrieval. MATERIALS AND METHODS: From October 2006 through June 2008, G2 IVC filters were placed in 140 consecutive patients who needed prophylaxis against pulmonary embolism (PE). General indications for filter placement included history of thromboembolic disease (n = 98) and high risk for PE (n = 42); specific indications included contraindication to anticoagulation (n = 120), prophylaxis in addition to anticoagulation (n = 16), and failure of anticoagulation (n = 4). Filter dwell time, technical success of filter retrieval, and complications related to placement or retrieval were retrospectively evaluated in patients who underwent filter removal. RESULTS: Twenty-seven attempts at G2 filter removal were made in 26 patients (12 men; age range, 24-88 years; mean age, 55.4 y) after a mean period of 122 days (range, 11-260 d). Data were collected retrospectively with institutional review board approval. Filter removal was successful in all 27 attempts (100%). Tilting of the filter (>= 15 degrees) occurred in five cases (18.5%), with probable filter incorporation into the right lateral wall of the IVC in one. Other complications of retrieval such as filter thrombosis, significant filter migration, filter fracture, and caval occlusion were not observed. CONCLUSIONS: G2 IVC filter retrieval has a high technical success rate and a low complication rate. Technical success appears to be unaffected by the dwell time within the reported range.
引用
收藏
页码:1046 / 1051
页数:6
相关论文
共 14 条
[1]   Initial experience in humans with a new retrievable inferior vena cava filter [J].
Asch, MR .
RADIOLOGY, 2002, 225 (03) :835-844
[2]   Retrievability of the recovery vena cava filter after dwell times longer than 180 days [J].
Binkert, CA ;
Sasadeusz, K ;
Stavropoulos, SW .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (02) :299-302
[3]   A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis [J].
Decousus, H ;
Leizorovicz, A ;
Parent, F ;
Page, Y ;
Tardy, B ;
Girard, P ;
Laporte, S ;
Faivre, R ;
Charbonnier, B ;
Barral, FG ;
Huet, Y ;
Simonneau, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 338 (07) :409-415
[4]   Experience with the recovery filter as a retrievable inferior vena cava filter [J].
Grande, WJ ;
Trerotola, SO ;
Reilly, PM ;
Clark, TWI ;
Soulen, MC ;
Patel, A ;
Shlansky-Goldberg, RD ;
Tuite, CM ;
Solomon, JA ;
Mondschein, JI ;
Fitzpatrick, MK ;
Stavropoulos, SW .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 16 (09) :1189-1193
[5]  
Grassi Clement J, 2003, J Vasc Interv Radiol, V14, pS271
[6]   Difficult retrieval of a recovery IVC filter [J].
Hagspiel, KD ;
Leung, DA ;
Aladdin, M ;
Spinosa, DJ ;
Matsumoto, AH ;
Angle, JF .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 15 (06) :645-647
[7]   Guidelines for the use of retrievable and convertible vena cava filters: Report from the Society of Interventional Radiology multidisciplinary consensus conference [J].
Kaufman, JA ;
Kinney, TB ;
Streiff, MB ;
Sing, RF ;
Proctor, MC ;
Becker, D ;
Cipolle, M ;
Comerota, AJ ;
Millward, SF ;
Rogers, FB ;
Sacks, D ;
Venbrux, AC .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (03) :449-459
[8]   IVC filter tilt and asymmetry: Comparison of the over-the-wire stainless-steel and titanium greenfield IVC filters [J].
Kinney, TB ;
Rose, SC ;
Weingarten, KE ;
Valji, K ;
Oglevie, SB ;
Roberts, AC .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1997, 8 (06) :1029-1037
[9]   Recovery G2 inferior vena cava filter: Technical success and safety of retrieval [J].
Oliva, Vincent L. ;
Perreault, Pierre ;
Giroux, Marie-France ;
Bouchard, Louis ;
Therasse, Eric ;
Soulez, Gilles .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 19 (06) :884-889
[10]   Five-year follow-up of prophylactic vena cava filters in high-risk trauma patients [J].
Rogers, FB ;
Strindberg, G ;
Shackford, SR ;
Osler, TM ;
Morris, CS ;
Ricci, MA ;
Najarian, KE ;
D'Agostino, R ;
Pilcher, DB .
ARCHIVES OF SURGERY, 1998, 133 (04) :406-411