IVC filter tilt and asymmetry: Comparison of the over-the-wire stainless-steel and titanium greenfield IVC filters

被引:36
作者
Kinney, TB
Rose, SC
Weingarten, KE
Valji, K
Oglevie, SB
Roberts, AC
机构
[1] UCSD Medical Center, Department of Radiology, San Diego, CA 92103-8756
关键词
embolism; pulmonary; thrombosis; venous; venae cavae; filters;
D O I
10.1016/S1051-0443(97)70706-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: A comparison of tilting, caval coverage, asymmetry, and insertion problems with the over-the-wire stainless-steel and titanium versions of the Greenfield filter. MATERIALS AND METHODS: The study compared 104 stainless-steel and 141 titanium Greenfield inferior vena cava (IVC) filter insertions, The angle the sheath and deployed filter made relative to the cava, as well as filter strut distribution, were determined from spot films, The proportionate caval coverage was computed from the cavogram (anteroposterior projection), Mean filter tilts, subgrouped by insertion site, and caval coverage were compared with the Student t test, whereas strut patterns were analyzed with a contingency table. RESULTS: The filter caval and sheath caval angles correlated, The filter caval angles varied with insertion site, but were lowest with a right jugular approach, Caval coverage was identical with both designs, The stainless-steel version resulted in a more uniform distribution of struts in comparison with the titanium version, The incidence of insertion problems was not significantly different between the filter types. CONCLUSIONS: While IVC filter tilting was not improved with the newer design, the pattern of struts was more uniformly symmetric with the stainless-steel device, The right jugular insertion site was associated with the lowest filter caval angles and the most symmetric pattern of struts.
引用
收藏
页码:1029 / 1037
页数:9
相关论文
共 20 条
[1]  
[Anonymous], 1965, Probability and statistics for engineers
[2]  
BROWN PP, 1974, CARDIOVASC SURG, V49, P272
[3]   Evaluation of a new percutaneous stainless steel Greenfield filter [J].
Cho, KJ ;
Greenfield, LJ ;
Proctor, MC ;
Hausmann, LA ;
Bonn, J ;
Dolmatch, BL ;
Eschelman, DJ ;
Flick, PA ;
Kinney, TB ;
Marx, MV ;
McFarland, DR ;
Ohki, SK ;
Pais, SO ;
Sussman, SK ;
Waltman, AC .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1997, 8 (02) :181-187
[4]  
CHO KJ, 1992, JVIR, V3, P40
[5]   PERCUTANEOUS INFERIOR VENA-CAVAL FILTERS - FOLLOW-UP OF 7 DESIGNS IN 320 PATIENTS [J].
FERRIS, EJ ;
MCCOWAN, TC ;
CARVER, DK ;
MCFARLAND, DR .
RADIOLOGY, 1993, 188 (03) :851-856
[6]  
GLANTZ SA, 1981, PRIMER BIOSTATISTICS, P30
[7]   EXPERIMENTAL EMBOLIC CAPTURE BY ASYMMETRIC GREENFIELD FILTERS [J].
GREENFIELD, LJ ;
PROCTOR, MC .
JOURNAL OF VASCULAR SURGERY, 1992, 16 (03) :436-444
[8]  
GREENFIELD LJ, 1988, SURGERY, V104, P706
[9]  
GREENFIELD LJ, 1983, SURG GYNECOL OBSTET, V156, P217
[10]  
GREENFIELD LJ, 1991, J VASC SURG, V14, P253