MR Elastography as a Method for the Assessment of Myocardial Stiffness: Comparison with an Established Pressure-Volume Model in a Left Ventricular Model of the Heart

被引:63
作者
Kolipaka, Arunark [1 ]
McGee, Kiaran P. [1 ]
Araoz, Philip A. [1 ]
Glaser, Kevin J. [1 ]
Manduca, Armando [1 ]
Romano, Anthony J. [2 ]
Ehman, Richard L. [1 ]
机构
[1] Mayo Clin, Rochester, MN 55905 USA
[2] USN, Res Lab, Washington, DC 20375 USA
基金
美国国家卫生研究院;
关键词
MRE; P-V relationships; gated MRE sequences; cardiac MRE; MAGNETIC-RESONANCE ELASTOGRAPHY; FAILURE; MUSCLE; ELASTICITY; MOTION; SENC;
D O I
10.1002/mrm.21991
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Magnetic resonance elastography (MRE) measurements of shear stiffness (mu) in a spherical phantom experiencing both static and cyclic pressure variations were compared to those derived from an established pressure-volume (P-V)-based model. A spherical phantom was constructed using a silicone rubber composite of 10 cm inner diameter and 1.3 cm thickness. A gradient echo MRE sequence was used to determine mu within the phantom at static and cyclic pressures ranging from 55 to 90 mmHg. Average values of mu using MRE were obtained within a region of interest and were compared to the P-V-derived estimates. Under both static and cyclic pressure conditions, the P-V- and MRE-based estimates of mu ranged from 98.2 to 155.1 kPa and 96.2 to 150.8 kPa, respectively. Correlation coefficients (R-2) of 0.98 and 0.97 between the P-V and MRE-based estimates of shear stiffness measurements were obtained. For both static and cyclic pressures, MRE-based measures of mu agree with those derived from a P-V model, suggesting that MRE can be used as a new, noninvasive method of assessing mu in sphere-like fluid-filled organs such as the heart. Magn Reson Med 62:135-140, 2009. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:135 / 140
页数:6
相关论文
共 28 条
[11]   MR elastography of breast cancer: Preliminary results. [J].
McKnight, AL ;
Kugel, JL ;
Rossman, PJ ;
Manduca, A ;
Hartmann, LC ;
Ehman, RL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 178 (06) :1411-1417
[12]   CLINICAL-ASSESSMENT OF DIASTOLIC FUNCTION [J].
MIRSKY, I ;
PASIPOULARIDES, A .
PROGRESS IN CARDIOVASCULAR DISEASES, 1990, 32 (04) :291-318
[13]   ASSESSMENT OF PASSIVE ELASTIC STIFFNESS FOR ISOLATED HEART-MUSCLE AND INTACT HEART [J].
MIRSKY, I ;
PARMLEY, WW .
CIRCULATION RESEARCH, 1973, 33 (02) :233-243
[14]   MAGNETIC-RESONANCE ELASTOGRAPHY BY DIRECT VISUALIZATION OF PROPAGATING ACOUSTIC STRAIN WAVES [J].
MUTHUPILLAI, R ;
LOMAS, DJ ;
ROSSMAN, PJ ;
GREENLEAF, JF ;
MANDUCA, A ;
EHMAN, RL .
SCIENCE, 1995, 269 (5232) :1854-1857
[15]   ELASTOGRAPHY - A QUANTITATIVE METHOD FOR IMAGING THE ELASTICITY OF BIOLOGICAL TISSUES [J].
OPHIR, J ;
CESPEDES, I ;
PONNEKANTI, H ;
YAZDI, Y ;
LI, X .
ULTRASONIC IMAGING, 1991, 13 (02) :111-134
[16]   Real-time imaging of regional myocardial function using Fast-SENC [J].
Pan, L ;
Stuber, M ;
Kraitchman, DL ;
Fritzges, DL ;
Gilson, WD ;
Osman, NF .
MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (02) :386-395
[17]  
Press W.H., 1992, NUMERICAL RECIPES FO, V2nd, P650
[18]  
RICKARDS AF, 1978, EUR J CARDIOL, V7, P213
[19]   Quantitative shear wave magnetic resonance elastography: Comparison to a dynamic shear material test [J].
Ringleb, SI ;
Chen, QS ;
Lake, DS ;
Manduca, A ;
Ehman, RL ;
An, KN .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (05) :1197-1201
[20]   MR elastography of the liver: Preliminary results [J].
Rouviere, Olivier ;
Yin, Meng ;
Dresner, M. Alex ;
Rossman, Phillip J. ;
Burgart, Lawrence J. ;
Fidler, Jeff L. ;
Ehman, Richard L. .
RADIOLOGY, 2006, 240 (02) :440-448