Comparative hemolysis study of clinically available centrifugal pumps

被引:28
作者
Naito, K
Suenaga, E
Cao, ZL
Suda, H
Ueno, T
Natsuaki, M
Itoh, T
机构
[1] Saga Medical School, Dept. Thorac. and Cardiovasc. Surg., Saga
[2] Dept. Thorac. and Cardiovasc. Surg., Saga 849, 5-1-1, Nabeshima
关键词
hemolysis; centrifugal pump; plasma-free hemoglobin; normalized index of hemolysis; pump rotational speed;
D O I
10.1111/j.1525-1594.1996.tb04480.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Centrifugal pumps have become important devices for cardiopulmonary bypass and circulatory assistance. Five types of centrifugal pumps are clinically available in Japan. To evaluate the blood trauma caused by centrifugal pumps, a comparative hemolysis study was performed under identical conditions. In vitro hemolysis test circuits were constructed to operate the BioMedicus BP-80 (Medtronic, BioMedicus), Sarns Delphin (Sarns/ 3M Healthcare), Isoflow (St. Jude Medical [SJM]), HPM-15 (Nikkiso), and Capiox CX-SP45 (Terumo). The hemolysis test loop consisted of two 1.5 m lengths of polyvinyl chloride tubing with a 3/8-inch internal diameter, a reservoir with a sampling port, and a pump head. All pumps were set to flow at 6 L/min against the total pressure head of 120 mm Hg. Experiments were conducted simulta neously for 6 h at room temperature (21 degrees C) with fresh bovine blood. Blood samples for plasma-free hemoglobin testing were taken, and the change in temperature at the pump outlet port was measured during the experiment. The mean pump rotational speeds were 1,570, 1,374, 1,438, 1,944, and 1,296 rpm, and the normalized indexes of hemolysis were 0.00070, 0.00745, 0.00096, 0.00066, 0.00090 g/100 L for the BP-80, Sarns, SJM, Nikkiso, and Terumo pumps, respectively. The change in temperature at the pump outlet port was the least for the Nikkiso pump (1.8 degrees C) and the most with the SJM pump (3.8 degrees C). This study showed that there is no relationship between the pump rotational speed (rpm) and the normalized index of hemolysis in 5 types of centrifugal pumps. The pump design and number of impellers could be more notable factors in blood damage.
引用
收藏
页码:560 / 563
页数:4
相关论文
共 9 条
[1]  
ARAKI K, 1994, JPN J ARTIF ORGANS, V23, P893
[2]   INVITRO ASSESSMENT OF CENTRIFUGAL PUMPS FOR VENTRICULAR ASSIST [J].
JAKOB, H ;
KUTSCHERA, Y ;
PALZER, B ;
PRELLWITZ, W ;
OELERT, H .
ARTIFICIAL ORGANS, 1990, 14 (04) :278-283
[3]  
KLETSCHKA HD, 1975, MINN MED, V58, P757
[4]  
MULLER MR, 1993, ARTIF ORGANS, V17, P103
[5]   THE NEED FOR STANDARDIZING THE INDEX OF HEMOLYSIS [J].
NAITO, K ;
MIZUGUCHI, K ;
NOSE, Y .
ARTIFICIAL ORGANS, 1994, 18 (01) :7-10
[6]  
NISHIDA H, 1993, NONPULSATILE PERFUSI, P24
[7]  
Noon G P, 1990, ASAIO Trans, V36, pM616
[8]   TOTAL ERYTHROCYTE DESTRUCTION TIME - THE NEW INDEX FOR THE HEMOLYTIC PERFORMANCE OF ROTARY BLOOD PUMPS [J].
SHIMONO, T ;
MAKINOUCHI, K ;
NOSE, Y .
ARTIFICIAL ORGANS, 1995, 19 (07) :571-575
[9]   A COMPACT CENTRIFUGAL BLOOD PUMP FOR EXTRACORPOREAL-CIRCULATION - DESIGN AND PERFORMANCE [J].
TANAKA, S ;
YAMAMOTO, S ;
YAMAKOSHI, K ;
KAMIYA, A .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1987, 109 (03) :272-278