Determination of trace Cd and Pb in environmental and biological samples by ETV-ICP-MS after single-drop microextraction

被引:134
作者
Li, Li [1 ]
Hu, Bin [1 ]
Xia, Linbo [1 ]
Jiang, Zucheng [1 ]
机构
[1] Wuhan Univ, Dept Chem, Wuhan 430072, Peoples R China
关键词
single-drop microextraction; electrothermal vaporization; ICP-MS; cadmium; lead; 8-hydroxyquinoline;
D O I
10.1016/j.talanta.2006.03.006
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method based on single-drop microextraction (SDME) combined with electrothermal vaporization (ETV)-ICP-MS was proposed for the determination of trace Cd and Ph. 8-Hydroxyquinoline (8-HQ) was employed as extractant dissolved in several microliters of chloroform and then an organic microdrop was formed at the tip of the microsyringe needle to extract the interest analytes. The vaporization behavior of the metal-8-HQ chelates in graphite furnace was investigated, and the ETV temperature program was optimized. The factors that influenced the extraction efficiency of target analytes (including pH value, flow rate of sample, extraction time and organic microdrop volume) were studied. Under the optimum conditions, the detection limits of the Cd and Ph were 4.6 and 2.9 pg mL(-1) with the enrichment factor of 140-fold for Cd and 190-fold for Ph, respectively. The proposed method was applied successfully to the determination of trace Cd and Ph in environmental and biological samples. In order to validate the developed method, a certified reference material of GBW 08501 peach leaves was analyzed and the determined values obtained were in a good agreement with the certified values. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:468 / 473
页数:6
相关论文
共 29 条
[1]  
Boiteau H., 1988, QUANTITATIVE TRACE A, P543
[2]   Determination of lead in water samples by graphite furnace atomic absorption spectrometry after cloud point extraction [J].
Chen, JR ;
Xiao, SM ;
Wu, XH ;
Fang, KM ;
Liu, WH .
TALANTA, 2005, 67 (05) :992-996
[3]  
Chen SZ, 1999, J ANAL ATOM SPECTROM, V14, P1723
[4]   Headspace single-drop microextration for the detection of organotin compounds [J].
Colombini, V ;
Bancon-Montigny, C ;
Yang, L ;
Maxwell, P ;
Sturgeon, RE ;
Mester, Z .
TALANTA, 2004, 63 (03) :555-560
[5]   Determination of cadmium, lead, iron, nickel and chromium in selected food matrices by plasma spectrometric techniques [J].
Cubadda, F ;
Raggi, A .
MICROCHEMICAL JOURNAL, 2005, 79 (1-2) :91-96
[6]  
DAHER RT, 1995, ANAL CHEM, V67, pR405
[7]  
GUO GQ, 1997, HDB ANAL CHEM, V1, P309
[8]   Molecular design of acyclic polyether dicarboxylic acids possessing pseudo-18-crown-6 frameworks for selective lead(II) extraction [J].
Hayashita, T ;
Sawano, H ;
Higuchi, T ;
Indo, M ;
Hiratani, K ;
Zhang, ZY ;
Bartsch, RA .
ANALYTICAL CHEMISTRY, 1999, 71 (04) :791-795
[9]   Liquid phase microextraction in a single drop of organic solvent by using a conventional microsyringe [J].
He, Y ;
Lee, HK .
ANALYTICAL CHEMISTRY, 1997, 69 (22) :4634-4640
[10]   Solvent microextraction into a single drop [J].
Jeannot, MA ;
Cantwell, FF .