Effect of surface finish on the osseointegration of laser-treated titanium alloy implants

被引:206
作者
Götz, HE
Müller, M
Emmel, A
Holzwarth, U
Erben, RG
Stangl, R
机构
[1] Univ Munich, Fac Vet, Inst Physiol Physiol Chem & Anim Nutr, Munich, Germany
[2] Friedrich Alexander Univ Hosp, Dept Surg, Trauma Surg Div, Erlangen, Germany
[3] Univ Appl Sci, Amberg, Germany
[4] P Brehm Co, Weisendorf, Germany
关键词
titanium alloy; laser; porosity; histomorphometry; osseointegration;
D O I
10.1016/j.biomaterials.2003.11.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
It was the purpose of this study to examine the osseointegration of laser-textured titanium alloy (Ti6Al4V) implants with pore sizes of 100, 200, and 300 pm, specifically comparing 200-mum implants with polished and corundum-blasted surfaces in a rabbit transcortical model. Using a distal and proximal implantation site in the distal femoral cortex, each animal received all four different implants in both femora. The bone-implant interface and the newly formed bone tissue within the pores and in peri-implant bone tissue were examined 3, 6, and 12 weeks post-implantation by static and dynamic histomorphometry. Here we show that additional surface blasting of laser-textured Ti6Al4V implants with 200-mum pores resulted in a profound improvement in osseointegration, 12 weeks postimplantation. Although lamellar bone formation was found in pores of all sizes, the amount of lamellar bone within pores was linearly related to pore size. In 100-mum pores, bone remodeling occurred with a pronounced time lag relative to larger pores. Implants with 300-mum pores showed a delayed osseointegration compared with 200-mum pores. We conclude that 200 mum may be the optimal pore size for laser-textured Ti6Al4V implants, and that laser treating in combination with surface blasting may be a very interesting technology for the structuring of implant surfaces. (C) 2003 Elsevier Ltd. All rights reserved.
引用
收藏
页码:4057 / 4064
页数:8
相关论文
共 34 条
[1]
[Anonymous], BIOMEDICAL ENG HDB
[2]
INTERMITTENT MICROMOTION INHIBITS BONE INGROWTH - TITANIUM IMPLANTS IN RABBITS [J].
ASPENBERG, P ;
GOODMAN, S ;
TOKSVIGLARSEN, S ;
RYD, L ;
ALBREKTSSON, T .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (02) :141-145
[3]
BOBYN JD, 1980, CLIN ORTHOP RELAT R, P263
[4]
OSSEOINTEGRATION AND ITS EXPERIMENTAL BACKGROUND [J].
BRANEMARK, PI .
JOURNAL OF PROSTHETIC DENTISTRY, 1983, 50 (03) :399-410
[5]
Brunski J B, 1999, Adv Dent Res, V13, P99
[6]
INFLUENCE OF SURFACE CHARACTERISTICS ON BONE INTEGRATION OF TITANIUM IMPLANTS - A HISTOMORPHOMETRIC STUDY IN MINIATURE PIGS [J].
BUSER, D ;
SCHENK, RK ;
STEINEMANN, S ;
FIORELLINI, JP ;
FOX, CH ;
STICH, H .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (07) :889-902
[7]
Influence of various structure treatments on histological fixation of titanium implants [J].
Chang, YS ;
Gu, HO ;
Kobayashi, M ;
Oka, M .
JOURNAL OF ARTHROPLASTY, 1998, 13 (07) :816-825
[8]
A role for surface topography in creating and maintaining bone at titanium endosseous implants [J].
Cooper, LF .
JOURNAL OF PROSTHETIC DENTISTRY, 2000, 84 (05) :522-534
[9]
Bone response to implant surface morphology [J].
D'Lima, DD ;
Lemperle, SM ;
Chen, PC ;
Holmes, RE ;
Colwell, CW .
JOURNAL OF ARTHROPLASTY, 1998, 13 (08) :928-934
[10]
Davies JE, 1998, INT J PROSTHODONT, V11, P391