Assessment of wear on the cones of modular stainless steel Exeter hip stems

被引:7
作者
Chaplin, RPS [1 ]
Lee, AJC [1 ]
Hooper, RM [1 ]
机构
[1] Univ Exeter, Dept Engn, Exeter EX4 4QF, Devon, England
关键词
D O I
10.1023/B:JMSM.0000042683.65855.6e
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The wear on the stem cones of retrieved Exeter Universal hip stems has been assessed using scanning electron microscopy (SEM), energy-dispersive X-ray analysis (EDX) and surface profilometry. The in-service life of these prosthetic stems varied, up to a maximum of 7 years. A combination of SEM, EDX and visual assessment indicates that the stem cones have not suffered from any corrosion. SEM scans indicate that damage to stem cones (excluding extraction and post-removal damage) can be categorised into insertion marks and fretting marks. In some cases there are signs of material being deposited on the cone surface. Surface profilometry suggests that the levels of debris generation at the cone/internal head interface are very low relative to those that are likely to be associated with head articulation against the acetabular cup. A total of 20 stem cones underwent SEM scans. From these, 10 subsequently have undergone surface profilometry along with the corresponding internal head surfaces. There is a good correlation between surface roughness measured by surface profilometry and the topography observed in the SEM images. The surface roughness of each stem cone is similar to that of the corresponding internal head surface. (C) 2004 Kluwer Academic Publishers.
引用
收藏
页码:977 / 990
页数:14
相关论文
共 31 条
[1]
Basle MF, 1996, J BIOMED MATER RES, V30, P157
[2]
BHAMBRI SK, 1994, AM SOC TEST MATER, V1173, P111, DOI 10.1520/STP18098S
[3]
BOGGAN RS, 1994, J LONG-TERM EFF MED, V4, P177
[4]
COLLIER JP, 1991, T SOC BIOMATERIALS, V14, P292
[5]
COOKE J, 1998, THESIS U EXETER
[6]
Effect of circumferential plasma-spray porous coating on the rate of femoral osteolysis after total hip arthroplasty [J].
Emerson, RH ;
Sanders, SB ;
Head, WC ;
Higgins, L .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (09) :1291-1298
[7]
FLEMMING CAC, 1994, AM SOC TEST MATER, V1173, P156, DOI 10.1520/STP18102S
[8]
IN-VIVO CORROSION OF MODULAR HIP-PROSTHESIS COMPONENTS IN MIXED AND SIMILAR METAL COMBINATIONS - THE EFFECT OF CREVICE, STRESS, MOTION, AND ALLOY COUPLING [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (12) :1533-1544
[9]
In vitro corrosion testing of modular hip tapers [J].
Goldberg, JR ;
Gilbert, JL .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 64B (02) :78-93
[10]
A comparative joint simulator study of the wear of metal-on-metal and alternative material combinations in hip replacements [J].
Goldsmith, AAJ ;
Dowson, D ;
Isaac, GH ;
Lancaster, JG .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H1) :39-47