Prefocal alignment improves stone comminution in shockwave lithotripsy

被引:25
作者
Sokolov, DL
Bailey, MR
Crum, LA
Blomgren, PM
Connors, BA
Evan, AP
机构
[1] Univ Washington, Appl Phys Lab, Ctr Ind & Med Ultrasound, Seattle, WA 98105 USA
[2] Indiana Univ, Sch Med, Dept Anat & Cell Biol, Indianapolis, IN USA
关键词
D O I
10.1089/08927790260472845
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background: The Dornier HM-3 machine continues to be one of the most effective lithotripters in use. However, tissue damage occurs in most, if not all, shockwave lithotripsy (SWL) treatments. Cavitation appears to contribute to desired stone comminution as well as to undesired tissue damage. Studies of cavitation in electrohydraulic shockwave lithotripters indicate that the greatest cavitation activity occurs, not at the geometric focus, F2, but at a site proximal to F2 by 1 to 3 cm. In clinical practice, however, stones are aligned with F2. Materials and Methods: In vitro stone comminution, hemolysis, and free-radical production were assessed. along the focal axis, and pig kidneys treated with SWL in vivo were sectioned to determine the extent of hemorrhagic injury along the focal axis. Model gypsum stones received 200 shockwaves in vitro at 18 kV. Results: At F2, the average number of fragments >1.5 mm was 1.3 +/- 0.5, and the weight loss was 11.3 +/- 1.1%. At 2 cm from F2 (F2-2 cm), these values increased to 4 +/- 2.8 and 16.1 +/- 4.2%, respectively. Samples of 10% hematocrit blood were similarly exposed. Hemolysis was equivalent at F2-2 cm (14.7 +/- 2.3%) and F2 (15.2 +/- 3%) but decreased significantly at all other positions. Samples of iodine solution received 1500 shockwaves at 20 kV. Hydroxyl radical production was greatest at F2-2 cm (0.384 +/- 0.035 muM) and decreased significantly distal to this position. The volume of tissue injury in pig kidneys was greatest with prefocal shockwave exposure. Conclusion: Stone comminution may be achieved more rapidly without greater tissue damage by a simple shift in stone alignment to F2-2 cm.
引用
收藏
页码:709 / 715
页数:7
相关论文
共 43 条
[1]  
AVERKIOU MA, 1996, NEW DEV MANAGEMENT U, P21
[2]   Comparison of electrohydraulic lithotripters with rigid and pressure-release ellipsoidal reflectors. I. Acoustic fields [J].
Bailey, MR ;
Blackstock, DT ;
Cleveland, RO ;
Crum, LA .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1998, 104 (04) :2517-2524
[3]   Comparison of electrohydraulic lithotripters with rigid and pressure-release ellipsoidal reflectors. II. Cavitation fields [J].
Bailey, MR ;
Blackstock, DT ;
Cleveland, RO ;
Crum, LA .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 106 (02) :1149-1160
[4]  
BAILEY MR, 2001, 17 INT C AC ROM IT
[5]  
Blomgren PM, 1997, ANAT REC, V249, P341, DOI 10.1002/(SICI)1097-0185(199711)249:3<341::AID-AR4>3.0.CO
[6]  
2-X
[7]   The search for cavitation in vivo [J].
Carstensen, EL ;
Gracewski, S ;
Dalecki, D .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (09) :1377-1385
[8]   In vivo pressure measurements of lithotripsy shock waves in pigs [J].
Cleveland, RO ;
Lifshitz, DA ;
Connors, BA ;
Evan, AP ;
Willis, LR ;
Crum, LA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (02) :293-306
[9]   Design and characterization of a research electrohydraulic lithotripter patterned after the Dornier HM3 [J].
Cleveland, RO ;
Bailey, MR ;
Fineberg, N ;
Hartenbaum, B ;
Lokhandwalla, M ;
McAteer, JA ;
Sturtevant, B .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (06) :2514-2525
[10]   PRESSURE WAVE-FORMS GENERATED BY A DORNIER EXTRACORPOREAL SHOCK-WAVE LITHOTRIPTER [J].
COLEMAN, AJ ;
SAUNDERS, JE ;
PRESTON, RC ;
BACON, DR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1987, 13 (10) :651-657