The effect of speed of sound in ultrasound test objects on lateral resolution

被引:22
作者
Dudley, NJ
Gibson, NM
Fleckney, MJ
Clark, PD
机构
[1] City Hosp Nottingham, Dept Phys Med, Nottingham NG5 1PB, England
[2] Newcastle Gen Hosp, Dept Phys Med, Newcastle Upon Tyne NE4 6BE, Tyne & Wear, England
关键词
speed of sound; resolution; ultrasound quality control; urethane-based test object; gel-based test object;
D O I
10.1016/S0301-5629(02)00648-8
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Urethane test objects are available that are more durable than traditional gel-based objects. Their ultrasound (US) propagation speed differs from scanner calibration velocity. This is clearly a disadvantage in checking caliper accuracy, but its effect on other measurements such as resolution is not well known. A test object containing a column of nylon filaments was constructed and filled with water at different temperatures to give US speeds of 1540 m s(-1) and 1450 m s(-1). Images of the filaments were digitally captured and analysed to obtain the full-width half-maximum of the target images at each depth. Target image width was greater at 1450 m s(-1). Measurements were also performed for the two types of tissue-mimicking test objects; results were in close agreement with the water test object. US beam profiles are broader in media with low propagation velocities. (E-mail: ndudley@ncht.trent.nhs.uk) (C) 2002 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:1561 / 1564
页数:4
相关论文
共 5 条
[1]   A computerised quality control testing system for B-mode ultrasound [J].
Gibson, NM ;
Dudley, NJ ;
Griffith, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (12) :1697-1711
[2]   The effect of acoustic velocity on phantom measurements [J].
Goldstein, L .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (07) :1133-1143
[3]   Liquid or solid ultrasonically tissue-mimicking materials with very low scatter [J].
Madsen, EL ;
Frank, GR ;
Dong, F .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (04) :535-542
[4]  
WELLS PNT, 1969, PHYSICAL PRINCIPLES
[5]  
Zagzebski JA, 1995, MED CT ULTRASOUND CU, P121