Accurate and reproducible target navigation with the Stereotaxis Niobe® Magnetic Navigation System

被引:49
作者
Armacost, Michelle P. [1 ]
Adair, Janet [1 ]
Munger, Torrey [1 ]
Viswanathan, Raju R. [1 ]
Creighton, Francis M. [1 ]
Curd, David T. [1 ]
Sehra, Ruchir [1 ]
机构
[1] Stereotaxis Inc, St Louis, MO 63108 USA
关键词
magnetic navigation; catheter ablation; arrhythmia; atrial fibrillation;
D O I
10.1111/j.1540-8167.2007.00708.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Accurate and Reproducible Target Navigation. Introduction: The Stereotaxis Niobe (R) Magnetic Navigation System (MNS, Stereotaxis, Inc., St. Louis, MO, USA) allows for remote-control navigation of magnetically enabled catheters and guidewires for clinical ablation and electrophysiology (EP) device placement using two permanent magnets located on opposite sides of the patient table. Our objective was to provide a clinical framework for expected navigational accuracy during a case by calibrating the system's reproducibility using a realistic heart phantom under various conditions. Methods and Results: We performed two sets of experiments to demonstrate the accuracy of magnetic catheters using the MNS. The first experiment calibrated deviations in the deflection of an EP catheter by the magnetic system from the expected angles using proprietary algorithms based on predicted geometry. We found that the magnetic fields produced catheter angulation movements within a mean of 4 degrees of biophysical predictions. The second experiment used the MNS to navigate these catheters to previously labeled target positions within a phantom heart model and estimated the actual displacement from desired target positions. We found that the accuracy for reaching the desired targets was 100%. Conclusions: Remote navigation of magnetically enabled EP catheters using the MNS accurately and reproducibly navigates to target sites in a heart phantom. This may enable more complex, successful, and time-efficient procedures in the cardiac catheterization laboratory.
引用
收藏
页码:S26 / S31
页数:6
相关论文
共 11 条
[1]   Outcomes after cardiac perforation during radiofrequency ablation of the atrium [J].
Bunch, TJ ;
Asirvatham, SJ ;
Friedman, PA ;
Monahan, KH ;
Munger, TM ;
Rea, RF ;
Sinak, LJ ;
Packer, DL .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2005, 16 (11) :1172-1179
[2]   Remote catheter ablation of parahisian accessory pathways using a novel magnetic navigation system - A report of two cases [J].
Ernst, S ;
Hachiya, H ;
Chun, JKR ;
Ouyang, F .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2005, 16 (06) :659-662
[3]   Initial experience with remote catheter ablation using a novel magnetic navigation system -: Magnetic remote catheter ablation [J].
Ernst, S ;
Ouyang, FF ;
Linder, C ;
Hertting, K ;
Stahl, F ;
Chun, J ;
Hachiya, H ;
Bänsch, D ;
Antz, M ;
Kuck, KH .
CIRCULATION, 2004, 109 (12) :1472-1475
[4]   Novel, magnetically guided catheter for endocardial mapping and radiofrequency catheter ablation [J].
Faddis, MN ;
Blume, W ;
Finney, J ;
Hall, A ;
Rauch, J ;
Sell, J ;
Bae, KT ;
Talcott, M ;
Lindsay, B .
CIRCULATION, 2002, 106 (23) :2980-2985
[5]   Magnetic guidance system for cardiac electrophysiology - A prospective trial of safety and efficacy in humans [J].
Faddis, MN ;
Chen, J ;
Osborn, J ;
Talcott, M ;
Cain, ME ;
Lindsay, BD .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 42 (11) :1952-1958
[6]   Heart-surface reconstruction and ECG electrodes localization using fluoroscopy, epipolar geometry and stereovision: Application to noninvasive imaging of cardiac electrical activity [J].
Ghanem, RN ;
Ramanathan, C ;
Jia, P ;
Rudy, Y .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (10) :1307-1318
[7]   Effects of operator experience on the outcome and duration of pulmonary vein isolation procedures for atrial fibrillation [J].
Knight, BP ;
Oral, H ;
Chugh, A ;
Scharf, C ;
Lai, SWK ;
Pelosi, F ;
Strickberger, SA ;
Morady, F .
AMERICAN JOURNAL OF CARDIOLOGY, 2003, 91 (06) :673-677
[8]  
LINDSAY BD, 2006, HEART RHYTHM, V3, pS125
[9]   Atrial fibrillation ablation: State of the art [J].
Pappone, C ;
Santinelli, V .
AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (12A) :59L-64L
[10]   Robotic magnetic navigation for atrial fibrillation ablation [J].
Pappone, C ;
Vicedomini, G ;
Manguso, F ;
Gugliotta, F ;
Mazzone, P ;
Gulletta, S ;
Sora, N ;
Sala, S ;
Marzi, A ;
Augello, G ;
Livolsi, L ;
Santagostino, A ;
Santinelli, V .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (07) :1390-1400