Measurements in radiotherapy beams using on-line mosfet detectors

被引:24
作者
Kron, T
Rosenfeld, A
Lerch, M
Bazley, S
机构
[1] Univ Wollongong, Ctt Med Radiat Phys, Wollongong, NSW 2522, Australia
[2] Univ Newcastle, Sch Math & Phys Sci, Callaghan, NSW 2308, Australia
[3] Newcastle Mater Misericordiae Hosp, Waratah, NSW 2298, Australia
关键词
D O I
10.1093/oxfordjournals.rpd.a006022
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
When acquiring data to characterise radiation beams for radiotherapy treatment planning measurements in steep dose gradients such as beam penumbra or dose build-up are often required. A metal oxide semiconductor field effect transistor (MOSFET) with its inherent high spatial resolution wits used for Penumbra measurements in a 120 kVp X ray beam. The customised MOSFET system features a Pulsed readout that allows the acquisition Of data points in User defined time intervals of less than 1 s. Using a modified scanning beam data acquisition system the penumbra was acquired on-line ill 0.1 mm steps. Measurements were made at different distances behind the beam collimator. From the extrapolation of the penumbra width to a location directly under the block the spatial resolution of file MOSFET system can be estimated to be better than 0.1 mm. This excellent spatial resolution has many potential applications in radiotherapy dosimetry. including the characterisation of multileaf collimator systems.
引用
收藏
页码:445 / 448
页数:4
相关论文
共 10 条
[1]   EVALUATION OF A PTW DIAMOND DETECTOR FOR ELECTRON-BEAM MEASUREMENTS [J].
HEYDARIAN, M ;
HOBAN, PW ;
BECKHAM, WA ;
BORCHARDT, IM ;
BEDDOE, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (08) :1035-1042
[2]   Improved spatial resolution by MOSFET dosimetry of an x-ray microbeam [J].
Kaplan, GI ;
Rosenfeld, AB ;
Allen, BJ ;
Booth, JT ;
Carolan, MG ;
Holmes-Siedle, A .
MEDICAL PHYSICS, 2000, 27 (01) :239-244
[3]   Dose response of various radiation detectors to synchrotron radiation [J].
Kron, T ;
Duggan, L ;
Smith, T ;
Rosenfeld, A ;
Butson, M ;
Kaplan, G ;
Howlett, S ;
Hyodo, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (11) :3235-3259
[4]   THE PENUMBRA OF A 6-MV X-RAY-BEAM AS MEASURED BY THERMOLUMINESCENT DOSIMETRY AND EVALUATED USING AN INVERSE-SQUARE ROOT FUNCTION [J].
KRON, T ;
ELLIOTT, A ;
METCALFE, P .
MEDICAL PHYSICS, 1993, 20 (05) :1429-1438
[5]   Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 1998, 25 (10) :1919-1927
[6]   GENERAL SPECIFICATIONS FOR SILICON SEMICONDUCTORS FOR USE IN RADIATION-DOSIMETRY [J].
RIKNER, G ;
GRUSELL, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (09) :1109-1117
[7]   MOSFET dosimetry of an X-ray microbeam. [J].
Rosenfeld, AB ;
Kaplan, GI ;
Kron, T ;
Allen, BJ ;
Dilmanian, A ;
Orion, I ;
Ren, B ;
Lerch, MLF ;
Holmes-Siedle, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (06) :1774-1780
[8]   Feasibility study of online high-spatial-resolution MOSFET dosimetry in static and pulsed X-ray radiation fields [J].
Rosenfeld, AB ;
Lerch, MLF ;
Kron, T ;
Brauer-Krisch, E ;
Bravin, A ;
Holmes-Siedle, A ;
Allen, BJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (06) :2061-2068
[9]  
*SIEM, 1993, MED ENG DAT FORM FAC
[10]   EVALUATION OF A DUAL BIAS DUAL METAL-OXIDE-SILICON SEMICONDUCTOR FIELD-EFFECT TRANSISTOR DETECTOR AS RADIATION DOSIMETER [J].
SOUBRA, M ;
CYGLER, J ;
MACKAY, G .
MEDICAL PHYSICS, 1994, 21 (04) :567-572