Inter-observer reliability of measurements performed on digital long-leg standing radiographs and assessment of validity compared to 3D CT-scan

被引:39
作者
Boonen, B. [1 ]
Kerens, B. [2 ]
Schotanus, M. G. M. [1 ]
Emans, P. [3 ]
Jong, B. [1 ]
Kort, N. P. [1 ]
机构
[1] Zuyderland Med Ctr, Dr H Vd Hoffplein 1, NL-6162 BG Sittard Geleen, Netherlands
[2] AZ Sint Maarten, Leopoldstr 2, B-2800 Mechelen, Belgium
[3] Maastricht Univ, Med Ctr, P Debyelaan 25, NL-6229 HX Maastricht, Netherlands
关键词
Long-leg radiographs; Reliability; Validity; 3D CT-scan; Total knee arthroplasty; TOTAL KNEE ARTHROPLASTY; LIMB ALIGNMENT; NAVIGATION; TKA;
D O I
10.1016/j.knee.2015.08.008
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Background: Long-leg radiographs (LLR) are often used in orthopaedics to assess limb alignment in patients undergoing total knee arthroplasty (11(A). However, there are still concerns about the adequacy of measurements performed on LLR. We assessed the reliability and validity of measurements on LLR using three-dimensional computed tomography (3D CT)-scan as a gold standard. Methods: Six different surgeons measured the mechanical axis and position of the femoral and tibial components individually on 24 LLR. Intradass correlation coefficients (ICC) were calculated to obtain reliability and Bland-Altman plots were constructed to assess agreement between measurements on LLR and measurements on 3D CT scan. Results: ICC agreement for the six observer measurements on LLR was 0.70 for the femoral component and 0.80 for the tibial component. The mean difference between measurements performed on LLR and 3D CT-scan was 0.3 degrees for the femoral component and - 1.1 degrees for the tibial component. Variation of the difference between LLR and 3D CT-scan for the femoral component was 1.1 degrees and 0.9 degrees for the tibial component. 95% of the differences between measurements performed on LLR and 3D CT-scan were between - 1.9 degrees and 2.4 degrees (femoral component) and between - 2.9 and 0.7 (tibial component). Conclusion: Measurements on LLR show moderate to good reliability and, when compared to 3D CT-scan, show good validity. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:20 / 24
页数:5
相关论文
共 20 条
[1]
The long leg radiograph is a reliable method of assessing alignment when compared to computer-assisted navigation and computer tomography [J].
Babazadeh, Sina ;
Dowsey, Michelle M. ;
Bingham, Roger J. ;
Ek, Eugene T. ;
Stoney, James D. ;
Choong, Peter F. M. .
KNEE, 2013, 20 (04) :242-249
[2]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]
Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial [J].
Boonen, B. ;
Schotanus, M. G. M. ;
Kerens, B. ;
van der Weegen, W. ;
van Drumpt, R. A. M. ;
Kort, N. P. .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2013, 21 (10) :2206-2212
[4]
Boonen B, 2015, KNEE SURG SPORT TR A, DOI [http://dx.doi.org/10.1007/500167-015-3661-1, DOI 10.1007/500167-015-3661-1]
[5]
Preliminary experience with the patient-specific templating total knee arthroplasty 40 cases compared with a matched control group [J].
Boonen, Bert ;
Schotanus, Martijn G. M. ;
Kort, Nanne P. .
ACTA ORTHOPAEDICA, 2012, 83 (04) :387-393
[6]
Plain Radiograph Fails to Reflect the Alignment and Advantages of Navigation in Total Knee Arthroplasty [J].
Choi, Won Chul ;
Lee, Sahnghoon ;
An, Joon Hwan ;
Kim, Dongwook ;
Seong, Sang Cheol ;
Lee, Myung Chul .
JOURNAL OF ARTHROPLASTY, 2011, 26 (05) :756-764
[7]
Agreement between radiological and computer navigation measurement of lower limb alignment [J].
Dexel, Julian ;
Kirschner, Stephan ;
Guenther, Klaus-Peter ;
Luetzner, Joerg .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2014, 22 (11) :2721-2727
[8]
Correlation of radiographic and navigated measurements of TKA limb alignment: a matter of time? [J].
Hauschild, Oliver ;
Konstantinidis, Lukas ;
Baumann, Tobias ;
Niemeyer, Philipp ;
Suedkamp, Norbert P. ;
Helwig, Peter .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2010, 18 (10) :1317-1322
[9]
The position and orientation of total knee replacement components A COMPARISON OF CONVENTIONAL RADIOGRAPHS, TRANSVERSE 2D-CT SLICES AND 3D-CT RECONSTRUCTION [J].
Hirschmann, M. T. ;
Konala, P. ;
Amsler, F. ;
Iranpour, F. ;
Friederich, N. F. ;
Cobb, J. P. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :629-633
[10]
Huang Teng-Le, 2004, J Chin Med Assoc, V67, P185