Automatic marker detection and 3D position reconstruction using cine EPID images for SBRT verification

被引:40
作者
Park, Sang-June [1 ]
Ionascu, Dan [1 ]
Hacker, Fred [1 ]
Mamon, Harvey [1 ]
Berbeco, Ross [1 ]
机构
[1] Harvard Univ, Sch Med, Dept Radiat Oncol, Dana Farber Brigham & Womens Canc Ctr, Boston, MA 02115 USA
关键词
automatic marker detection; 3D position reconstruction; cine EPID; SBRT verification; IMPLANTED FIDUCIAL MARKERS; SCALE SELECTION; EDGE-DETECTION; RADIOTHERAPY; TRACKING; KV; MV; FEASIBILITY; ALGORITHM; MODE;
D O I
10.1118/1.3218845
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
In previous studies, an electronic portal imaging device (EPID) in cine mode was used for validating respiratory gating and stereotactic body radiation therapy (SBRT) by tracking implanted fiducials. The manual marker tracking methods that were used were time and labor intensive, limiting the utility of the validation. The authors have developed an automatic algorithm to quickly and accurately extract the markers in EPID images and reconstruct their 3D positions. Studies have been performed with gold fiducials placed in solid water and dynamic thorax phantoms. In addition, the authors have examined the cases of five patients being treated under an SBRT protocol for hepatic metastases. For each case, a sequence of images was created by collecting the exit radiation using the EPID. The markers were detected and recognized using an image processing algorithm based on the Laplacian of Gaussian function. To reduce false marker detection, a marker registration technique was applied using image intensity as well as the geometric spatial transformations between the reference marker positions produced from the projection of 3D CT images and the estimated marker positions. An average marker position in 3D was reconstructed by backprojecting, towards the source, the position of each marker on the 2D image plane. From the static phantom study, spatial accuracies of <1 mm were achieved in both 2D and 3D marker locations. From the dynamic phantom study, using only the Laplacian of the Gaussian algorithm, the marker detection success rate was 88.8%. However, adding a marker registration technique which utilizes prior CT information, the detection success rate was increased to 100%. From the SBRT patient study, intrafractional tumor motion (3.1-11.3 mm) in the SI direction was measured using the 2D images. The interfractional patient setup errors (0.1-12.7 mm) in the SI, AP, and LR directions were obtained from the average marker locations reconstructed in 3D and compared to the reference planning CT image. The authors have developed an automatic algorithm to extract marker locations from MV images and have evaluated its performance. The measured intrafractional tumor motion and the interfractional daily patient setup error can be used for off-line retrospective verification of SBRT. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3218845]
引用
收藏
页码:4536 / 4546
页数:11
相关论文
共 17 条
[1]
Towards fluoroscopic respiratory gating for lung tumours without radiopaque markers [J].
Berbeco, RI ;
Mostafavi, H ;
Sharp, GC ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (19) :4481-4490
[2]
A technique for respiratory-gated radiotherapy treatment verification with an EPID in cine mode [J].
Berbeco, RI ;
Neicu, T ;
Rietzel, E ;
Chen, GTY ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (16) :3669-3679
[3]
A novel method for estimating SBRT delivered dose with beam's-eye-view images [J].
Berbeco, Ross I. ;
Hacker, Fred ;
Zatwarnicki, Chris ;
Park, Sang-June ;
Ionascu, Dan ;
O'Farrell, Desmond ;
Mamon, Harvey J. .
MEDICAL PHYSICS, 2008, 35 (07) :3225-3231
[4]
Clinical feasibility of using an epid in cine mode for image-guided verification of stereotactic body radiotherapy [J].
Berbeco, Ross I. ;
Hacker, Fred ;
Ionascu, Dan ;
Mamon, Harvey J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (01) :258-266
[5]
Feature tracking with automatic selection of spatial scales [J].
Bretzner, L ;
Lindeberg, T .
COMPUTER VISION AND IMAGE UNDERSTANDING, 1998, 71 (03) :385-392
[7]
Robust fluoroscopic respiratory gating for lung cancer radiotherapy without implanted fiducial markers [J].
Cui, Ying ;
Dy, Jennifer G. ;
Sharp, Greg C. ;
Alexander, Brian ;
Jiang, Steve B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (03) :741-755
[8]
On the use of EPID-based implanted marker tracking for 4D radiotherapy [J].
Keall, PJ ;
Todor, AD ;
Vedam, SS ;
Bartee, CL ;
Siebers, JV ;
Kini, VR ;
Mohan, R .
MEDICAL PHYSICS, 2004, 31 (12) :3492-3499
[9]
DETECTING SALIENT BLOB-LIKE IMAGE STRUCTURES AND THEIR SCALES WITH A SCALE-SPACE PRIMAL SKETCH - A METHOD FOR FOCUS-OF-ATTENTION [J].
LINDEBERG, T .
INTERNATIONAL JOURNAL OF COMPUTER VISION, 1993, 11 (03) :283-318
[10]
Edge detection and ridge detection with automatic scale selection [J].
Lindeberg, T .
INTERNATIONAL JOURNAL OF COMPUTER VISION, 1998, 30 (02) :117-154